Humidity probe calibration (435-2/-4 only)
This function is only available if a humidity probe is plugged in.
The calibration values can be reset to the default settings (Reset). A 2-point
calibration can be performed.
²
Resetting tthe ccalibration vvalues:
The configuration menu is open, config. is displayed.
1 Probe ¬¬Calibr. ¬ .
2 Press / to select Reset and confirm by pressing twice.
- The calibration values are reset to the default settings.
²
Calibrating:
The configuration menu is open, config. is displayed.
1 Probe ¬¬Calibr. ¬ .
2 Press / to select calibration point P1 or P2 and confirm by
pressing twice.
3 Put the humidity probe into the reference medium and wait for the
equalisation period to elapse.
- The current humidity reading and the calibration point (nominal value) are
displayed.
4 Start the calibration menu with .
5 Save calibration with or cancel calibration with .
P internal (435-3/-4 only)
The internal pressure sensor can be switched on/off.
²
Switching tthe iinternal ppressure ssensor oon/off:
The configuration menu is open, config. is displayed.
1 Device ¬¬P intern ¬ .
2 Press / On or Off and confirm with .