EasyManua.ls Logo

TESTO 435 Series - Parameter

TESTO 435 Series
64 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
²
Activating/deactivating rreading ccalculation:
The main menu is open, Menu is displayed.
1 Calc. ¬ .
2 Press / to select the variable that is to be activated/deactivated
and confirm with .
3 Press / to select On (= activated) or Off (= deactivated) and
confirm with .
²
Activating ddifferential ccalculation ((delta):
The instrument is in measurement view.
Differential calculation is performed with the parameters that are shown on
the display.
1 Press and to select the measurement channels from which the
difference is to be calculated.
2 Press to open the main menu.
3 calc. ¬ .
4 Press / to select Delta and confirm with .
6.2.6 Parameter
Some calculated variables relate to particular reference values (ambient
conditions or factors for certain probes). These can be entered by means of
calculation parameters.
Parameters used for calculating variables:
Parameters Calculated measuring variables
Temp. (reference temperature) Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
Humid. (reference humidity) Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
Pres. (reference pressure) Flow velocity, volumetric flow rate (for measurement with a
pitot tube or hot wire probe)
Area (cross-section area) Volumetric flow rate
P-factor (pitot tube factor) Flow velocity and volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
OK
OK
OK
OK
OK
6. Setting the instrument54

Related product manuals