7 Using the product
26
3
Press [▲] / [▼] to select Pressure
Leak Test and press [Menu/Enter]
to confirm.
-
For the temperature-compensated leak test, a connection is
established with available Bluetooth
®
probes. If cable probes are
connected to the instrument, they are prioritized for the compensation.
Please note that only air probes are ideal for temperature-
compensated leakage testing.
-
testo 915i / testo 605i is switched on and automatically connected.
Other temperature probes compatible with testo 570s can be
connected.
-
The Pressure Leak Test menu is displayed.
T Comp is shown on the display if a compatible probe is connected via
Bluetooth
®
or cable. The temperature compensation is used for the
measurement result.
4
Press the [▼] (Start) key.
-
The leakage test is carried out.
5
Press the [▼] (Stop) key.
-
The leakage test is terminated.
-
The measurement result is
displayed.