17
i
Activate/deactivate laser, continuous measurement and illumination.
i Enter headers for printer.
Measurement configuration
Limit values (Factory setting)
Infrared Dew point distance Surface moisture
upper limit value 950.0 - -20.0
lower limit value 10.0 -40.0 -
i Activate/deactivate
Audible alarm
Reset
i Activate/deactivate
Allow store deletion
- Delete memory.
Memory contents are deleted.
- Factory settings
Instrument settings are returned to factory settings
Humidity module adjustment
i Press
Adjustment humidity module
button.
The window
Dialog
opens.
i To adjust press the buttons 11.3% and 75.3%.
Close window
7.5.2 Ending programming
i Click on
Apply
to apply the programming carried out to testo 845.
7.6 Disconnecting
i In the window
Archive
, click on the connection you want to disconnect
with the right-hand mouse key.
i Select
Close
.
The connection to testo 845 is disconnected.
7. Programming