Shenzhen TestPad Electronic Co., Ltd. http://www.test-pad.com
16
14.
OVC-Max:
This
parameter
can
be
adjusted
only
when
the
over-current
test
is
enabled;
it
can be increased or decreased by 0.1A, each time, and the maximum value is 12A.
15.
R1-Min:
This
parameter
can
be
adjusted
only
when
the
R1
test
is
enabled;
it
can
be
increased or decreased by 0.1K, each time, and the maximum value is 999.9K.
16.
R1-Max:
This
parameter
can
be
adjusted
only
when
the
R1
test
is
enabled;
it
can
be
increased or decreased by 0.1K, each time, and the maximum value is 999.9K.
17.
R2-Max:
This
parameter
can
be
adjusted
only
when
the
R2
test
is
enabled;
it
can
be
increased or decreased by 0.1K, each time, and the maximum value is 999.9K.
18.
R2-Min:
This
parameter
can
be
adjusted
only
when
the
R2
test
is
enabled;
it
can
be
increased or decreased by 0.1K, each time, and the maximum value is 999.9K.
19.
CHECK
ID:
Select
MOTO
(indicating
that
the
ID
of
an
MOTO
mobile
phone
can
be
checked) or No (indicating that no ID is tested).
20. ID Type: Select one from 10 ID types and one self-learning ID type.
21.
Short
TEST:
For
Li-ion
battery,
enable
or
disable
the
test.
For
other
battery
types,
this
function is unavailable.
22.
Short
TIME:
It
can
be
adjusted
in
the
range
of
1mS~200mS.
If
the
tested
battery
does
not
cut off the current within the specified period, it is deemed that the battery short-circuit test
fails.
23.
Short
Re:
If
Charge
is
selected,
it
indicates
that
the
battery
output
is
restored
through
temporary charge. If OPEN is selected, it indicates that the battery output is restored
through load disconnection. If the battery voltage is not restored to the lower threshold, it is
deemed that the short-circuit test fails.
24.
OPEN
time:
It
is
valid
if
Short
Re
is
set
to
OPEN.
It
indicates
that
the
entire
load
will
be
disconnected after the short-circuit protection and then the voltage will be tested after the
time set here.
25. detV_Max (0.5C): The maximum value is 1V.
26.
Exit
Mode:
If
Save
is
selected,
the
settings
after
each
adjustment
will
be
saved
in
the
internal memory and will not be lost after the device power-off. If NoSave is selected, the
settings will not be saved and will be lost after the device power-off. If Reset is selected,
the default settings will be restored after you exit the menu.
After all the parameters are set, press ENTER. Then the tester will automatically start and follow
the process below to test all the selected items: