B
·
R
·
A
·
H
·
M
·
S GmbH
B
·
R
·
A
·
H
·
M
·
S KRYPTOR compact PLUS User Manual (Version R08en) Page 17 of 30
B
.
R
.
A
.
H
.
M
.
S KRYPTOR compact SOLUTIONS 1 to 4
For SOLUTIONS 1 and 2, the dead volume has been defined to finish a reconstitution started even if red level is
detec t ed; it is normal to have around 12 mL in SOLUTION 1 and approximately 8 mL in SOLUTION 2 when red
status appears on the interface screen. For SOLUTIONS 3 and 4 the dead volume is around 4 mL, defined to finish
wash step.
COLOUR
EXPLANATION
Green: OK
Yellow: Take care to change the bottle soon / solution expires tomorrow
Red: Bottle to be replaced = dead volume reached or solution expired
Fluidic System
An intermediate tank system enables you to refill B
.
R
.
A
.
H
.
M
.
S KRYPTOR BUFFER or distilled/demineralized water
bottle when red status is detected without stopping the current run by use of dead volume in intermediate tank
system. If the dead volume is used then waste red level is detected and the instrument stops at the end of current
run or prime. A prime is mandatory after refilling a bottle.
COLOUR EXPLANATION
Green: OK
Red: Waste bottle full (void), buffer or water bottle empty (filling), intermediate waste collector full (Call Hot Line)

Related product manuals