27
Rejects
Path: Rejects
Clear Reject Counter
This button clears the reject counter displayed on the top menu. Press to reset to ‘0’.
Test Reject
This button when pressed will activate the reject output. This is used to test the action of the reject device. This will not inhibit
normal occurring rejects and the test function will return to normal when leaving this menu (Exit).
Setup Timing/Basic Setup
Delay
The reject delay time is the time delay between a detection and the activation of the reject device. This menu is used to change
that time delay. Use the Plus and Minus keys to change the value; test the change, then press Accept to accept this value or
Exit to restore the old value.
Duration
The reject duration is the length of time for which the reject device is activated after a detection has occurred. Use the Plus and
Minus keys to change the value; test the change then press Accept to accept this value or Exit to restore old value.
Learn
This menu is used to do a one-step reject timing calibration which is the easiest way to set the reject parameters.
Learn Procedure
1. Place a detectable metal sample in the center of the product. Pass the product through the detector. (do not
use large metal such as coins, etc.)
2. Press the Reject key when you wish to activate the reject device and hold the key down for the desired
duration.
3. The delay, duration and reject confirm (if used) time will all be calculated. Use individual menus to make any
small adjustments required to assure clean, complete rejection of a contaminated product.