Why does library batch import fail in Thermo Scientific Ion Torrent Genexus Integrated Sequencer?
- AAmber SimpsonAug 18, 2025
If the library batch import fails in your Thermo Scientific Laboratory Equipment, this may be because one or more entries in the library batch import spreadsheet contains special characters, line breaks, unexpected spaces, incorrect entry length, incorrect date formatting, or other formatting errors. Check each entry for correct formatting, correct any errors, and repeat the import. This issue may also arise if blank rows were copied into the library batch import template file from a different source. Rows that appear empty can contain hidden formatting that conflicts with the import function, so start with a clean library batch import template file, and be careful to copy only those rows that contain actual data.