Dionex ICS-5000 Ion Chromatography
System Operator's Manual
Document No. 065342
Revision 06
December 2011

Table of Contents

Questions and Answers

  • J
    Johnny GreerSep 8, 2025
    What to do if system backpressure is below 14 MPa in Thermo Scientific Dionex ICS-5000 Laboratory Equipment?
    • B
      Barbara FitzgeraldSep 9, 2025
      If the system backpressure is below 14 MPa, install a backpressure coil in the EG after the RFIC Eluent Degasser.
  • J
    Jenna LopezAug 23, 2025
    What to do if there's a flow system leak ahead of the cell in Thermo Scientific Dionex ICS-5000?
    • B
      Brenda BennettAug 23, 2025
      If you observe a flow system leak ahead of the cell, check all fittings and liquid lines for leaks. Tighten or replace all liquid line connections as necessary.
  • C
    christinafarmerAug 25, 2025
    How to address low system backpressure due to a loose fitting in Thermo Scientific Laboratory Equipment?
    • B
      balexanderAug 25, 2025
      To resolve low system backpressure caused by a loose fitting, ensure there are no liquid leaks in the flow system. Inspect tubing connections throughout the system, including check valves, injection valves, and columns, and tighten or replace fittings as needed.
  • M
    michael39Aug 26, 2025
    What to do about high system backpressure due to restriction in Thermo Scientific Dionex ICS-5000?
    • C
      Christopher BeckerAug 26, 2025
      If there is high system backpressure due to a restriction in the system plumbing, work backward through the system, starting at the flow cell exit. Loosen each fitting one at a time and observe the pressure.
  • G
    gmillerAug 27, 2025
    What to do if Thermo Scientific Dionex ICS-5000 has insufficient system equilibration?
    • J
      Jessica EricksonAug 28, 2025
      If the system has insufficient equilibration, especially after changes to operating parameters, allow a longer system equilibration time (up to 2 hours) before starting operation.
  • M
    Michael LawrenceSep 22, 2025
    What to do if Thermo Scientific Dionex ICS-5000 Laboratory Equipment's TC temperature stabilizer is blocked?
    • J
      Jeffery BakerSep 22, 2025
      If the TC temperature stabilizer is blocked, rinse the temperature stabilizer and the temperature stabilizer inlet line, using an appropriate solvent.
  • E
    Elaine WigginsSep 16, 2025
    What to do if Thermo Scientific Dionex ICS-5000 Laboratory Equipment's RFIC Eluent Degasser tubing is ruptured?
    • M
      Maria BrewerSep 16, 2025
      If the RFIC Eluent Degasser tubing is ruptured, replace the degasser.
  • J
    Jerry LoganSep 18, 2025
    What to do if Thermo Scientific Dionex ICS-5000 has an internal EGC leak?
    • S
      sandrastevensSep 18, 2025
      If there is an internal EGC leak, replace the EGC.
  • W
    Wyatt BlairSep 7, 2025
    What to do for inadequate system backpressure in Thermo Scientific Dionex ICS-5000 Laboratory Equipment?
    • J
      jackramseySep 7, 2025
      To address inadequate system or cell backpressure, add backpressure tubing to the cell outlet or to the pump (if no column is installed).
  • A
    April BrockSep 4, 2025
    What to do if Thermo Scientific Laboratory Equipment DP/SP is not properly primed?
    • R
      Rachel ArmstrongSep 4, 2025
      If the DP/SP is not properly primed, prime the pump.

Thermo Scientific Dionex ICS-5000 Specifications

General IconGeneral
TypeIon Chromatography System
Data SystemChromeleon Chromatography Data System
Pressure RangeUp to 5000 psi
Power Requirements100-240 V, 50/60 Hz
DetectorConductivity
PumpDual-Piston Pump
Column Temperature Range5°C to 80°C
ApplicationEnvironmental, Pharmaceutical
Flow Rate Range0.01 to 5.0 mL/min
Eluent GenerationOptional Eluent Generation Cartridge
DimensionsVaries with configuration
WeightVaries with configuration