What to do if Thermo Scientific Laboratory Equipment's piston seal is damaged?
- AAndrea ScottSep 6, 2025
If the piston seal is damaged, it allows liquid leaks that are usually visible. If necessary, replace the piston seal.
What to do if Thermo Scientific Laboratory Equipment's piston seal is damaged?
If the piston seal is damaged, it allows liquid leaks that are usually visible. If necessary, replace the piston seal.
What to do for inadequate system backpressure in Thermo Scientific Dionex ICS-5000 Laboratory Equipment?
To address inadequate system or cell backpressure, add backpressure tubing to the cell outlet or to the pump (if no column is installed).
What to do if system backpressure is below 14 MPa in Thermo Scientific Dionex ICS-5000 Laboratory Equipment?
If the system backpressure is below 14 MPa, install a backpressure coil in the EG after the RFIC Eluent Degasser.
How to fix noisy baseline with regular oscillation in Thermo Scientific Dionex ICS-5000?
If you observe regular baseline oscillation on high-sensitivity ranges, reconnect the short length of titanium tubing to the cell inlet.
What causes high background from CR-TC contamination in Thermo Scientific Dionex ICS-5000?
If there is a high background due to CR-TC contamination, clean the trap column as instructed in the CR-TC manual.
Why is there high background due to wrong eluent in Thermo Scientific Laboratory Equipment?
If there is high background due to wrong eluent, check that you are using the correct eluent for your application.
Type | Ion Chromatography System |
---|---|
Data System | Chromeleon Chromatography Data System |
Pressure Range | Up to 5000 psi |
Power Requirements | 100-240 V, 50/60 Hz |
Detector | Conductivity |
Pump | Dual-Piston Pump |
Column Temperature Range | 5°C to 80°C |
Application | Environmental, Pharmaceutical |
Flow Rate Range | 0.01 to 5.0 mL/min |
Eluent Generation | Optional Eluent Generation Cartridge |
Dimensions | Varies with configuration |
Weight | Varies with configuration |