User Manual
MaxQ 2000
Open-Air Platform Shaker
Operating Manual 057-278-00 (7004310) Rev. 13

Questions and Answers

  • N
    Nicole ZimmermanAug 1, 2025
    Why is my Thermo Scientific Laboratory Equipment shaker not working?
    • N
      nelsonalexandraAug 1, 2025
      If your Thermo Scientific Laboratory Equipment shaker isn't operating, start by ensuring the power cord is securely plugged in. If it is, verify that the power supply matches the requirements listed on the data label, and use a matching power supply. For E-Class models, check the fuse and replace it if needed. Also, for E-Class, look for flashing lights on the control panel and press any switch. If the elapsed timer is flashing, reset it. For A-Class models, make sure the timer isn't in the 'off' position; set it for continuous or timed operation. Finally, for A-Class, check the power switch and replace it if it's defective.
  • M
    Mr. Andre WhiteAug 15, 2025
    What to do if the platform of my Thermo Scientific MaxQ 2000 doesn't rotate properly?
    • H
      Heather StevensAug 15, 2025
      If the platform of your Thermo Scientific Laboratory Equipment doesn't rotate or has an erratic speed, check the power to the motor and replace the motor if defective. Inspect the drive belt and replace it if it's worn, broken, or has slipped off the pulley. For A-Class models, verify the power to the speed control and replace it if defective.

Summary

Section 1 Safety Information

Safety Warnings and Precautions

General safety precautions and warnings for operating the shaker to prevent hazards like electrical shock and personal injury.

Section 2 General Specifications

A-Class Control Panel Features

Details the features and operation of the A-Class control panel, including power switch, speed control, and timer functions.

E-Class Control Panel Features

Details the features and operation of the E-Class control panel, including speed display, timer, and membrane switches.

Section 4 Operation

Turning the Shaker On

Procedure for powering the shaker on and off using the main power switch.

Setting Shaking Speed

Instructions for adjusting and setting the desired shaking speed using the control panel.

Calibrating Shaking Speed

Steps for calibrating the shaker's displayed speed against a calibrated handheld tachometer.

Thermo Scientific MaxQ 2000 Specifications

General IconGeneral
Chamber MaterialStainless steel
ControllerMicroprocessor-based
DisplayLED
TypeShaker Incubator
Temperature Uniformity±0.5°C at 37°C
Orbit Diameter19 mm (0.75 in.)
Capacity2 x 1000 mL flasks
TimerUp to 99 hours 59 minutes
Voltage120V or 230V
Weight59 kg (130 lbs)