6.3 Preprogrammed
Settings
Settings can be customized via the Settings and Alerts menus.
Temperature Display & Alarm – The display is preset to read the glycol sensor
but can be changed to read the air sensor. Temperatures are detected and
displayed in Celsius units to one tenth of a degree. For example, 4.0°C. Upper
and lower alarm thresholds are driven by door the sensor selected for the display.
• Set Point Temperature – The unit preset/default is set at 5.0°C (39.2°F)
• Upper Threshold Alarm – The unit preset/default is set at 8.0°C
(46.4°F)
• Lower Threshold Alarm – The unit preset/default is set at 2.0°C
(35.6°F)
6.4 Preset alarms/alerts
The following alarms/alerts are factory preset:
6.5 Set Point
Temperature
When you first turn on the unit, the fan will run and the refrigerator will work to
reach the default set point temperature. Once set point is reached, the fan speed
will reduce to conserve energy. The unit will reach Air Temperature setting
rapidly. Reaching Glycol set point temperature will take longer.
When the unit is first plugged in and powered, the Temperature Range Alarms
are not enabled until the unit reaches its preset temperature range + an offset of