Navigating TSX 5.5 Cu.Ft. Refrigerator Controls
30 Installation and Operation Thermo Scientific TSX 5.5 cu.ft Undercounter Refrigerator
8.7.4.3 Calibration
When Calibrating, it is recommended to Calibrate the Control sensor first and
allow time for the system to equilibrate before making any adjustments to
Glycol. An example of calculating an offset value: Your NIST certified
temperature reading is 3.7°C, the unit reading is 4.2°C, the offset would be
-0.5°C.