EasyManua.ls Logo

Thermo iCAP 6000 Series - Page 11

Thermo iCAP 6000 Series
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iCAP 6000 Series ICP-OES Spectrometer Hardware Manual
Page 11 of 51 Version 1.5 – March 2006
The quartz body of the torch should be pushed into the metal torch mount. Ensure the torch
body is pushed fully into the metal torch mount. The marked line on the quartz torch is used to
locate the radial view hole correctly and should touch the metal body.
Correct Incorrect
Figure 2 Torch Assembly – torch body
Figure 3 Torch Assembly – centre tube
Insert the centre tube (ground glass joint) fully into the plastic centre tube holder.

Related product manuals