The quartz body of the torch should be pushed into the metal torch mount. Ensure the torch
body is pushed fully into the metal torch mount. The marked line on the quartz torch is used to
locate the radial view hole correctly and should touch the metal body.
Correct Incorrect
Figure 2 Torch Assembly – torch body
Figure 3 Torch Assembly – centre tube
Insert the centre tube (ground glass joint) fully into the plastic centre tube holder.