v. Higher limit torque (Default: 0)
If the value is set less than lower limit torque the display shows an error "SETERR".
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).
vi. Trigger torque (Default: 0)
It is the setting for starting angle measurement. If angle feature is OFF, this setting is
skipped.
If the value is set out of tool range the display shows an error "SETERR".
If it is set to "0", the double tightening detection angle, Hi/ Lo limit angle settings are
skipped and changed to "0". Then the display goes to tightening direction setting.
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).
vii. Double tightening detection angle (Default: 0)
Set the angle for double tightening detection angle judgment.
If measurement mode was set to “MODE-M”, this setting is skipped.
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).