ー34ー
6-2. Sterilizing-warming function
The autoclave allows a sterilizing-warming operation which is convenient for sterilizing agar
medium an similar materials.
After sterilization, a warming operation is carried out in order to avoid coagulation.
To change the time display form, refer to “6-9. Setting the time display form”.
To change the warming time, refer to “6-8. Setting the warming time”.
Set operating conditions following the procedures
given below.
Press the [Sterilize-warming] key.
The sterilizing-warming indication lamp lights
and the sterilizing-warming course is selected.
Concurrently, the temperature raising process
indication lamp, temperature and pressure
raising process indication lamp, sterilizing
process indication lamp, temperature and
pressure reduction process indication lamp,
warming process indication lamp, and
temperature reduction process indication lamp
on the work monitor light in yellow.
Press the [Temperature setting] keys to set the temperature for the sterilizing process.
Press the [Time setting] keys to set the time for the sterilizing process.
Press the [Warming/timer setting] keys to set the temperature for the warming process.
To change the warming time, refer to “6-8. Setting the warming time”.
Press the [Exhaust speed setting] keys to set the exhaust speed for the temperature and
pressure reduction process and temperature reduction process.
Press the [Cooling fan setting] key to set the number of fans activated during the temperature
and pressure reduction process and temperature reduction process.
The time display of the warming process is changed in units of one hour.
<
Setting sterilizing-warming operation
>
Sterilizing process Warming process
Settable temperature
range
Settable time range
Settable temperature
range
Settable time range
SX-300E/500E/700E 105
135℃ 0 : 01
9 : 59 45
95℃ 1 to 99
hours