Hold down the [Function] key when the ready indication lamp is illuminated.
After approximately two seconds, “F01” blinks at the temperature display.
Display “F04” by pressing the [Temperature setting] keys.
The code of the display form is shown at the time display.
Press the [Start] key.
The time display blinks and the code of the display form is displayed.
Press the [Time setting] keys to display the code of the desired form.
①
②
③
④
6-9. Setting the time display form
The display form of the time display can be changed.
Press the [Start] key to determine the code of the display form.
※
The code is set to “0 (hour: minute)” at the time of shipment from the factory.
⑤
Hold down the [Function] key when the ready indication lamp is illuminated.
After approximately two seconds, “F01” blinks at the temperature display.
Display “F03” by pressing the [Temperature setting] keys.
The code of the exhaust setting mode for liquid sterilization operation is displayed at the
time display.
Press the [Start] key.
The time display blinks.
Press the [Time setting] keys to display the code of the desired exhaust setting mode.
①
②
③
④
Regarding exhaust for liquid sterilization operation, the variable setting permitting ne
adjustment of valve opening can be selected as well as the normal exhaust speed setting.
Press the [Start] key to determine the code of the exhaust mode for liquid sterilization
operation.
※
The code is set to “0 (exhaust speed setting)” at the time of shipment from the factory.
⑤
<
Memorizing the preferred course
>
Code Display form Display range
0 Hour : minute 0 : 01
〜
9 : 59
1 Hour 1 to 99 hours
2 Minute 1 to 999 minutes
3 Hour : minute / hour 0 : 01
〜
9 : 59 / 10 to 99 hours
6-10.
Setting the exhaust mode for liquid sterilization operation
Code Exhaust setting mode Contents
0 Exhaust speed setting Six-step setting
1 Variable setting Variable