EasyManua.ls Logo

Tomy SX-700E - 6-3.Heating-Warming Function; 6-4.Memory Function

Tomy SX-700E
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ー35ー
6-3. Heating-warming function
The autoclave allows a heating-warming operation which is convenient for such usage as
dissolving agar.
After heating, a warming operation is carried out in order to avoid coagulation.
To change the time display form, refer to “6-9. Setting the time display form”.
To change the warming time, refer to “6-8. Setting the warming time”.
Set operating conditions following the procedures
given below.
Press the [Dissolve-warming] key.
The heating-warming indication lamp lights
and the heating-warming course is selected.
Concurrently, the temperature raising process
indication lamp, heating process indication
lamp, warming process indication lamp, and
temperature reduction process indication lamp
on the work monitor light in yellow.
Press the [Temperature setting] keys to set the temperature for the heating process.
Press the [Time setting] keys to set the time for the heating process.
Press the [Warming/timer setting] keys to set the temperature for the warming process.
To change the warming time, refer to “6-8. Setting the warming time”.
Press the [Exhaust speed setting] keys to set the exhaust speed for the temperature and
pressure reduction process and temperature reduction process.
Press the [Cooling fan setting] key to set the number of fans activated during the temperature
and pressure reduction process and temperature reduction process.
When “−−” is set by reducing the temperature for the heating process, the heating process only can
be cancelled.
When “−−” is set by reducing the temperature for the warming process, the warming process only
can be cancelled.
The time display of the warming process is changed in units of one hour.
<
Setting heating-warming operation
>
Set and display preferred operating conditions.
Hold down the [Memory] key.
The buzzer sounds approximately after two seconds and preferred operating conditions are
memorized.
All operating conditions displayed are memorized.
<
Memorizing the preferred course
>
6-4. Memory function
Preferred operating conditions can be called up easily by memorizing them.
Heating process Warming process
Settable temperature
range
Settable time range
Settable temperature
range
Settable time range
SX-300E/500E/700E 45
104℃ 0 : 01
9 : 59 45
95℃ 1 to 99
hours

Other manuals for Tomy SX-700E