9-2. Error code table
1
To ensure safety operation, the Autoclave employs the lid open/close detection system. To activate the system please open
or close the lid every time before the operation is started or the timer is set.
To resume operation after a blackout without opening the lid, please contact your dealer or our local service representative.
Error code Cause Countermeasures
E01
Sterilizing water is insufcient. Replenish sterilizing water.
The water level sensor is contaminated. Clean the water level sensor.
Deionized water was used in the chamber. Drain deionized water and add tap water.
(Refer to 5-5. Checking sterilizing water.)
Extremely cold water was used in the chamber. Add hot water and raise the temperature.
E04
The volume of sterilizing water in the chamber
becomes insufcient during operation.
Turn OFF the power supply switch temporarily. Open
the chamber lid after the pressure in the chamber falls to
0 MPa and replenish sterilizing water.
E05
The temperature in the chamber is abnormally
higher than the set value during sterilization
operation, heating operation, or warming.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E06
The pressure rises because the form of articles
to be sterilized is likely to retain air.
Turn OFF the power supply switch temporarily. Open
the chamber lid after the pressure in the chamber falls to
0 MPa. Take countermeasures so that air can escape
easily from articles or reduce the volume of articles.
(Refer to 5-6. Placing articles to be sterilized.)
E07
The temperature sensor has fractures. Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E08
The temperature in the chamber is abnormally
below the set value during sterilization
operation, heating operation, or warming.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E09
The chamber lid is open. Close the lid securely.
When this error occurs during operation, unplug the
cable from the outlet and contact your dealer or our local
service representative.
The lid open/close detection system
is not working.
1
Open or close the lid.
E10
The interlock of the chamber lid is not
activated.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E11
The temperature does not rise during the
temperature raising process or the temperature
and pressure raising process.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E12
The pressure sensor of articles to be sterilized
has fractures or is defective.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E13
The temperature sensor of articles to be
sterilized has fractures or is defective.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E15
The valve through which air escapes is
defective.
Unplug the power plug of the autoclave from the outlet
and contact the dealer or business institution of our
company.
E18
Lid parts are at the end of their service life and
need to be replaced.
Contact the dealer or business institution of our
company.
(Press the STOP key to temporarily cancel the error and
return to the operational mode.)