5 Turn the unit on again and allow auto-leveling to
complete.
6 Make a new mark (Mb) where the laser beam
strikes the paper.
7 Measure the distance between the rst mark (Ma)
and the second mark (Mb). No calibration is neces-
sary if distance is within 6mm (.23 inches).
8 Repeat procedure for the Y axis.
(2) Adjusting Calibration
If the distance between either set of marks is more than 6mm (.23 inches) but less than
25mm (1 inch), turn the unit off by pressing the [START] pad once and using the following
procedure. Conrm that unit has shut off before beginning the procedure.
(In steps 2 and 3, use of optional RC-30 remote control can be helpful. See page 34.)
30m (98feet)
X1 laser beam
X2 laser beam
If less than 6mm(.23 inches)
No calibration Necessary