Topcon RL-VH3C - Vertical Calibration - Beam Split Measurement

Topcon RL-VH3C
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
7. Without moving the position of the front foot, pivot the instrument so the rotary head is
now facing wall B.
8 Mark where the split beam emitted from
the top of the rotary head strikes wall B
(Hb). Measure the distance (dHb)
between marks Mb and Hb.
9 Compare the two measurements dHa and
dHb. If the difference between the two
measurements is less than 4mm (5/32 of
an inch), no adjustment is necessary. Oth-
erwise, adjust as follows.
about 1m