Page 25
5.7 Inert Gas Supply Connection for Model DCC
An inert gas supply is required for process operation of the Model DCC. The selection of the approved gases
listed below is primarily dependant upon compatibility with the materials being processed.
Warning! This oven is designed for use with only the non-explosive, non-flammable, inert gases
listed below. Operator injury and equipment damage may result if these instructions are not
followed.
The following inert gases are approved for the Models DCI and DCC.
Argon
Carbon Dioxide
Helium
Nitrogen
The inlet connection is located at the rear right side of the oven. A gas Flowmeter is installed at the front of
the oven with a Run Flow Adjustment needle valve.
Caution! A manually operated shut-off valve and a reliable pressure regulator should be installed
between your inert gas supply source and the oven connection.
The gas supply should be regulated by the customer in the range from 40 to 70 PSIG maximum. The
connection is Type 3/8” FPT. Make sure the connection is secure.
Note: The size and / or SCFH rating of the gas flowmeter will increase with larger ovens.
INERT GAS SUPPLY CONNECTION
GAS FLOWMETER
Rear of Oven
Front of Oven
Run Flow
Adjustment