EasyManua.ls Logo

Transcell Technology TI-1500B - Key-In Zero Calibration Value (F19)

Transcell Technology TI-1500B
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. While in the Setup mode, scroll to F19, then scroll down once using the ZERO ( )
key.
2. The display will momentarily show ET C0, followed by a flashing zero. Use the four
directional keys to adjust the displayed value to the zero calibration value.
3. After setting the exact value, press the PEAK (SET) key to save the value. The display
will showSETmomentarily, then revert back up to F19.
6.6 Key-in Zero Calibration Value (F19)
Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid span calibration value, obtained from a successful F17
calibration procedure, must be used.
1. While in the Setup mode, scroll to F20, then scroll down once using the ZERO ( )
key. The indicator will prompt you to enter the information in the follow table.
2. If the value shown is correct, press PEAK (SET) key to save the value, then move to
the next value. Otherwise, use the four directional keys to adjust the actual calibration
value.
3. After setting the exact value, press the PEAK (SET) key to save the value.
4. If the entered values are entered successful, the display will show SETmomentarily
before continuing to the next parameter. At the completion of the sequence, the indicator
will then revert back up to F20.
Name
Value
Zero calibration value
First test weight value
First span calibration value
Second test weight value
Second span calibration value
Third test weight value
Third span calibration value
Recommend: Please record 6.4,6.5,6.6 value into the table immediately and print
the table. In the case of non-volatile memory loss, you will key-in the value.

Related product manuals