19
18
TROUBLESHOOTING
SPECIFICATIONS
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
massage
Display will not
light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Display is dim or
show
Batteries are low.
E 1 shows The cuff is not secure.
Refasten the cuff and then
measure again.
E 2 shows The cuff is very tight
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E10 or E11
shows
E20 shows
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again
EExx,shows on
the display.
A calibration error
occurred (XX can be
some numeric
characters,such as 01,
02 and so on .
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
E21 shows
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
Readjust the cuff ,not too
loose or too tight and then
measure again.
The monitor detected
motion,talking or the
pluse is too poor
while measuring.
Relax for a
moment and then
measure again.
About 22~32cm or 22cm~42cm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Temperature:-20℃-60℃
Relative Humidity 10%-93%
Atmospheric Pressure: 50-106 kPa
Temperature:5℃ to 40℃ Relative humidity ≤85%
Atmospheric pressure: 86kPa to 106kPa
Power supply
Battery powered mode:
6VDC 4×AAA batteries
AC adaptor powered mode:
(Please only use the recommended AC
adaptor model).
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Net Weight
Approx.250g(Excluding the dry cells)
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
IP21
Accuracy
Normal working condition
Storage & transportation
condition
Software Version V01
Pressure:
5℃-40℃within±0.4kpa(3mmHg)
pulse value:±5%
Rated cuff pressure:
0kpa - 40kpa (0mmHg~300mmHg)
Measurement pressure: 5.3kPa-30.7kPa
(40mmHg-230mmHg)
pulse value: (40-199) beat/minute
Digital LCD V.A.60×92mm
Approx.140×130×49.7mm
4×AAA batteries,user manual
6V 1A