This manual contains relevant guidance on
routine maintenance; however, you may wish
to introduce additional checks as required by
OSHA. The following table shows the frequency
of these checks.
Operation Frequency
Clean Internally and Externally Daily
Visual Inspection Daily
Charge Battery Pack/Packs Daily
Functional Test* Daily, Weekly and Monthly
* Remarks: Dust from wood, grain and cement
may require more frequent replacement of the
filter. No filter should be left in the hood/helmet
for more than 1 month if in daily use or 3 months
if used infrequently.
Routine Checks
The expected life of the visor is approximately 3
years if well maintained and cleaned. (Care must
be taken not to scratch the visor).
Storage
Q The AIR/PRO should be placed in the carry
bag, stored within a temperature range of
41°F to 104°F and out of direct sunlight.
Q Replacement filters are to be left in air sealed
bags and expiry dates to be checked
regularly.
Transportation
Q The complete AIR/PRO should be
transported in the carry bag provided. Good
quality commercial packing should be used.
Symptom Cause Remedy
Reduced Cover air holes blocked Clear air flows
Filters clogged Fit a new filter
Battery flat Recharge battery or fit recharged
battery pack
No airflow Unit switched OFF Press switch to ON
Switch faulty Replace switch housing
Battery pack not fitted Fit a fully charged battery pack
Battery pack discharged Fit a fully charged battery pack
Fan-motor unit failure Fit a replacement fan-motor unit
Airflow