Appendix C-4
Gauge Appears to be Unstable
Perform a stability (stat) test (20 - 1 minute counts).
Perform a drift test (5 - 4 minute counts).
If the above tests pass, recheck the calibration parameters
and samples.
Gauge will not Communicate with Printer
Check the cable for continuity.
Check baud rates.
Check cable pin-out (see page B-4).
Make sure all other parameters match:
o Data bits = 8
o Stop bits = 2
o Parity = none
o Protocol = DSR/DTR
Software is not Functioning Properly
Re-initialize the gauge.
Contact customer service.