Troxler 3241-C - Drift Test

Troxler 3241-C
131 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-4
D
D
r
r
i
i
f
f
t
t
T
T
e
e
s
s
t
t
Assuming that the stat test has already been performed (and
passed) but gauge readings seem to drift between tests, the Drift
Test can be performed to check the long-term drift of the gauge.
The Drift Test consists of five counts performed approximately 3-
4 hours after completion of a stat test.
NOTE
The count time is based on the last stat test. The
drift test count time should be four times the stat
test count time. The gauge should not be turned off
between the stat test and drift test. The stat test
must be current.
Press SHIFT and 9 to display the Special menu. To access the
Drift Test feature, press
2 and ENTER for:
Remove any samples or sample pans from the gauge and press
START.
After completion of the 5 counts, the gauge will display the
average counts and indicate if the test passed or failed. The
Pass/Fail criterion is 1.0%.
Wi l l t ake f i ve
xx
mi n.
count s.
Cl ose
door
and
Pr ess
START
Avg Cnt s: xxxx
Dr i f t :
x. xx%