Model 3241 Series
4-11
How many sampl es
( 2- 12) ?
____
I nput
and
Pr ess
ENTER
- Sampl e #1 -
%AC:
_____%
I nput
and
Pr ess
ENTER
Fi t coef f : x. xxx
Want
t o
r evi ew
i nput
dat a?
The gauge will request the number of calibration samples.
Input the number of samples and press YES. The display will
request the %AC of the first sample.
Input the %AC, from line G. Press ENTER.
Place the first calibration sample in the sample chamber and
press
START. Record the counts on line M.
Repeat the above procedure for each sample. After all samples
have been measured the display will be:
The gauge calculates the calibration curve and curve “fit.” The
“fit coefficient” is any number between 0.0 and 1.0, with 1.0
being a perfect fit.
After calculation of the fit coefficient, the data may be viewed.
To view data, press YES at the above display.
If you do not want to review the data, press NO at the
prompt.