6–8
Place the gauge on the reference standard block. Make sure the source rod
is in the SAFE position and press START/ENTER. The display will be:
- DRIFT TEST -
Read #x Time:xxx
After the gauge completes five 4-minute counts, the display will be:
DENS DRIFT TEST
Avg cnts: xx
Drift: x.x % P
-ENTER for Moist
To view the moisture data from the last drift test, press START/ENTER.
The display is:
MOIST DRIFT TEST
Avg cnts: xx
Drift: x.x % P
View Drift data?
VIEWING DRIFT TEST DATA
The data for each of the readings may be viewed on the screen or may be
printed. This function is the same as viewing the stat test data.
Refer to page 6–5 for instructions on viewing or printing drift test data.
Figure 6–2 shows a sample drift test printout.
********************************
DRIFT TEST
********************************
Gauge Serial #: 123456
O6/15/2007 3:01 PM
READING | COUNTS
# | DENSITY MOISTURE
--------------------------------
1 2758 269
2 2749 270
3 2757 267
4 2759 271
5 2759 268
Average: 2756 269
Drift: 0.5 % 0.4 %
(PASS) (PASS)
Figure 6–2. Drift Test Printout Sample