6–12
SPECIAL CALIBRATION
The Special Calibration function allows the operator to re-calibrate the
gauge constants temporarily for use in measuring particular materials.
These materials may be limestone or other materials that do not fall within
the range of a normal calibration. Once enabled, the Special Calibration
will remain in effect until the gauge is turned off or Normal calibration is
re-enabled.
To execute Special Calibration, press SHIFT and SPECIAL for:
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
Press YES two times for the display:
YES- Next menu
6-SPECIAL CALIB.
7- NOMOGRAPH
8- PRECISION
Press 6 for the display:
SPEC CALIB.-OFF
Press: 1- enable
2- disable
3- Recalib.
At this time, the operator may either enable an existing special calibration
or disable an existing calibration that has been in use. The operator may
also re-calibrate the gauge for use in a different material.
ENABLE SPECIAL CALIBRATION
Press 1 to enable the calibration. The display will be:
SPECIAL SOIL
CALIBRATION
ENABLED!
The gauge will return to the READY display.

Table of Contents

Other manuals for Troxler 3401

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3401 and is the answer not in the manual?

Troxler 3401 Specifications

General IconGeneral
BrandTroxler
Model3401
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals