EasyManuals Logo

Troxler NTO 4730 User Manual

Troxler NTO 4730
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
6–8
CORRECTION MENU
The oven can store up to twenty aggregate correction factors
(ACFs). Each ACF is stored using an alphanumeric ID entered by
the operator. The ACF ID can be up to twelve characters. When a
correction factor is active, all measurements are adjusted using that
correction factor.
The Correction Menu enables the operator to select (activate) an
existing ACF, to create a new ACF, or to erase one or more stored
ACFs. To access the Correction Menu, press 4 at the MAIN
MENU. The oven displays:
Correction Menu
1. Select ACF
2. Create ACF
3. Erase ACF
To select an item from the menu, press the corresponding number
key.
SELECT ACF
To select (activate) an existing ACF, press 1 at the Correction
Menu shown on page 6–8. If there are no stored ACFs, the oven
displays the error mess
age No ACFs Found and returns to the
Correction Menu.
If there are stored ACFs, the oven displays:
Select
1. Scroll ACFs
2. Enter ACF ID

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler NTO 4730 and is the answer not in the manual?

Troxler NTO 4730 Specifications

General IconGeneral
BrandTroxler
ModelNTO 4730
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals