Core Modules 13
NOTICE - Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON while there is a leakage
in the sample line, this may reduce lifetime of devices.
Never remove collection vessels from the 4x Air Drying Unit. Drain off condensed water via
collection vessel tubing only.
After you have discarded condensed water of separating bottle, carefully reconnect bottle to
tubing.
Select "Status/Calo" and check for proper "SampleFlow” on "Calo Current Values" dialog (see
Software Operating Instructions - Calorimetry, Section "Status Menu, Calo").
The test area should be limited by a door or similar so no rapid alteration in the CO2/O2
concentration of the reference air can take place during the measurement.
A warmed-up calorimetry system should be used for an experiment only. The warming-up
period should be at least 2 hours.
Exactly specified gases with a test certificate should be used for calorimetric calibration only.
Calibration gases (O2, CO2) which concentration is known accurately to two decimal places
should be available.
The calorimetry sensors should be calibrated under the same conditions that are to be used
for the measurement (e.g. room temperature, pressure, location).
The calorimetry devices of the system should be leak-free, as otherwise the measurements
will be inaccurate.
The setting of the flowmeter (Calorimetry Master Control Unit) for the measurement should be
the same as for the calibration process.
For better comparability of test data, the room conditions, particularly temperature and
humidity, should be kept as constant as possible.
Air pumps (push principle respirometry, overpressure principle)
o The pump is designed for pumping air only. The performance values can be found on
the pump label.
o Ensure that the operating air temperature is within the range -10...+40°C (14…104°F)
o Ensure that humidity is <90%.
Vacuum pumps (pull principle respirometry, underpressure principle)
o The pump is suitable exclusively for air and NOT for pumping liquids or gases. Max.
permissible humidity <90%.
o The ambient temperature must not fall below -C (23°F) and must not exceed +40°C
(104°F). You are NOT allowed to operate the pump close to explosive liquids or on
potentially explosive premises.
o Unusual noises may indicate that a diaphragm has broken. In this case, unplug device
from mains supply and contact TSE Service
.
When the High-Speed Sensor Unit is in operating mode, ensure that the Oxymat sensor is
supplied with reference air (scavenging air) to protect the device against contamination.
NOTICE - Reduced lifetime!
If High-Speed Sensor Unit is powered ON without operating the system, contamination may
damage the sensors.
During operating mode of High-Speed Sensor Unit, ensure that…
o …its sensors are supplied with air from reference gas cylinder continuously.
o …all other devices of the CaloSys module entered operating mode as well.
o …PhenoMaster software is running.
Power OFF High-Speed Sensor Unit after use.