1: Introduction
2
Controller
module
Low Flow
module
High Flow
module
Oxygen
sensor
Figure 1. The Certifier
®
FA Test System
WARNING
To avoid the risk of explosion, do not use in the presence of
flammable anesthetic gases.
Only qualified and trained service technicians are authorized to
service the Certifier
®
FA Test System. Use only factory-approved
parts and procedures to service the device.
CAUTION
To avoid inaccurate test readings, do not obstruct tubing or inlet
or exhaust ports, and always use dry gas.
To avoid damage to the Certifier
®
FA Test System components,
always use bacteria filters upstream of the flow modules, and
always cap flow module ports when not in use.
TSI flowmeters are not medical devices under FDA 510(k) and in
no situation should be used for human measurements.