31
6 Specifications
NOTE: Specifications are subject to change without notice.
6.1 Physical
Controller module: 13.2 cm x 7 cm x 3.3 cm (5.2 in. x 2.8 in. x 1.3 in.).
High Flow module: 18 cm x 6.6 cm x 4 cm (7.1 in. x 2.6 in. x 1.6 in.).
Low Flow module: 12.7 cm x 5.1 cm x 2.8 cm (5.0 in. x 2.0 in. x 1.1 in.).
flow inlet: 22-mm female ISO taper.
flow outlet: 22-mm male ISO taper.
Low Flow module:
flow inlet: 0.25-in. barb.
flow outlet: 0.25-in. barb.
Approximately 1.36 kg (3 lb) for the standard kits.
Operating: 5 to 40 C (41 to 104 F). 15 to 80% relative humidity from
5 to 31 C decreasing linearly to 15 to 50% relative humidity at 40 C.
Storage: -40 to 70 C (-40 to 158 F) at 10 to 95% relative humidity.
Operating: 57.1 to 106 kPa (8.28 to 15.37 psia).
Storage: 50 to 106 kPa (7.25 to 15.37 psia).
Indoor Use
Operating Altitude up to 4000 m (13,000 ft)
Pollution degree I or II
Four commercially available AA batteries.
6.4 Test Measurements (see notes at end of section)
Air and oxygen: 2% of reading
or 0.075 SLPM, whichever is
greater.
Air/oxygen mixtures: 4% of
reading or 0.1 SLPM, whichever
is greater.
Air and oxygen: 2% of
reading or 0.010 SLPM,
whichever is greater.
Nitrous oxide: 4% of reading
or 0.025 SLPM, whichever is
greater.