Technical Section 
Calibration 
 
 
The EverWatch Face Velocity Monitor is a very versatile device which will work with a variety of 
fume hoods. Because fume hoods vary in design and performance, the EverWatch monitor must 
be calibrated on the fume hood it is installed. The calibration procedure is quite simple and can 
be accomplished in a matter of minutes with a little preparation and practice. 
 
Every EverWatch monitor shipped has an orange WARNING label attached indicating calibration 
is required: 
 
WARNING: 
Unit is not calibrated. 
Remove label only after field 
calibration is complete. 
 
This label is to be removed only after field calibration is complete. 
 
Before beginning to calibrate the EverWatch monitor, take a few minutes to review the calibration 
procedures below. The following equipment is required to perform the calibration: 
  Tape 
  Portable Air Velocity Meter such as TSI VELOCICALC
®
 or VELOCICHECK
® 
meters 
 
The calibration procedure can be broken down into the following: 
  Entering the calibration menu. 
  Establishing a zero flow velocity. 
  Setting the velocity calibration span. 
  Exiting the calibration menu. 
 
Calibration Procedure 
Power up the EverWatch monitor and allow the unit to warm up (30 to 60 seconds) before 
beginning the calibration procedure. On initial power up the keypad is disabled and SEnS is 
indicated on the display for approximately 30 seconds. 
 
Note:  All menu items described below are found in the Calibration menu. 
 
1.  Select CAL (calibration menu). See “Software Programming” if not familiar with 
keystroke procedure. 
 
2.  Place a piece of tape over the velocity sensor opening to seal off the air flow through 
the sensor (off white colored sensor is located inside of fume hood). 
 
3.  Select LCAL (zero calibration). Press the NORMAL key and CAL will flash for two 
minutes. During this time the monitor automatically calibrates the sensor for zero 
flow. 
 
Note:  If COdE is flashing on the display, the access code enable function is set to 
on and an access code is required to enter the calibration menu. Enter the 
calibration menu access code (code is located on last page of manual).