The SASH CAL OPEN item is used to record sash position 
of the fume hood. This is determined in one of two ways: 
1.  If a Vertical Sash Position Sensor is being used, this 
menu item will record the physical position of the 
Potentiometer with the sash open. 
2.  If there is no Vertical Sash Position sensor, then the 
controller will use both the face velocity and flow 
information in the sash open position to calibrate the 
sash % open equation. 
Upon entering this menu item, the display will read “Raise 
Sash, then Press Enter.” Make sure the sash is raised, and 
then press the Enter key. 
This menu item will timeout after a period of one hour if the 
Enter key is not pressed. No information will be saved if 
timeout occurs. 
This menu item will not appear if the controller is not 
configured to accept a sash sensor or flow measurement 
with sidewall sensor.