>> Ultrasonic Flow meter User Manual
This is not a window but a command to print the pipe parameters.
By default, the produced data will be directed to the internal serial bus (thermal printer).
You can also direct those data to the serial communication port.
This is not a window but a command to print the diagnostic information.
By default, the produced data will be directed to the internal serial bus (thermal printer).
You can also direct those data to the serial communication port.
This is not a window but a command to copy the current display window. By default, the
produced data will be directed to the internal serial bus (thermal printer). You can also
direct those data to the serial communication port.
By use of the window copying function, you can hardcopy very window displaying
manually by switching windows, or you can obtain the window displaying data by
communication.
Browse the 32 recorded instrument power-on and power-off date and time with the flow
rate at the time of power on and off
Displays the total working time of the flow meter.
When the backup battery is removed, the total working time will be reset to zero.
Displays the last power-off date and time
Displays the last power-off flow rate
Displays how many times of has been powered on and powered off.
A scientific calculator for the convenience of field working.
All the values are in single accuracy.
The calculator can be used while the flow meter is conducting flow measurement.
Water density and PT100 temperature can also be found in this function.
Set fluid sound speed threshold
Whenever the estimated sound speed (displayed in M92) exceeds this threshold, an
alarms signal will be generated and can transmitted to BUZZER or OCT or RELAY.
This function can used to produce an alarm or output when fluid material changes.
Displays total flow for this month(only for the time past)
Displays total flow for this year(only for the time past)
Display the not-working total time in seconds. The total failure timer will also include
the time when power off, if the back-up battery is applied.
Entry to solidify the zero point. Password protected.
Setup the Q value threshold.
If the present Q is below this threshold, flow rate will be set to 0.
This function is useful when flow meter is installed in noisy environment or on airy
pipes.
The maximum flow rates for today and this month.
Serial port tester with CMM command output for very second.
Entry to hardware adjusting windows only for the manufacturer
4mA calibration for AI3 input
20mA calibration for AI3 input
4mA calibration for AI4 input