>> Ultrasonic Flow meter User Manual
22
M84
Set the thermal energy unit:
0. GJ 1. KC 2.KWh 3. BTU
M85
Select temperature sources
0. from T1,T2 (factory default)
1. from AI3,AI4
M86
Select the Specific Heat Value.
Factory default is ‘GB’. Under this setting, the flow meter will calculate the enthalpy of
water based on the international standard.
If the fluid is other than water, you should select option ‘1. Fixed Specific Heat’, and
enter the specific heat value of the fluid.
M87
Turn on or turn off the Energy totalizer.
M88
Select thermal energy totalizer multiplying factor.
Factory default is ‘1’.
M89
1. Display the temperature difference
2. Window for entering the lowest temperature difference.
M8.
Heat meter is on
1. Inlet
2. Outlet
Select the heat meter installation place.
M90
Display signal strengths S (one for upstream and one for downstream), and signal quality
Q value.
Signal strength is presented by 00.0 to 99.9, the bigger the value, the bigger the signal
strength will be, and more reliable readings will be made.
Q value is presented by 00 to 99, the bigger the better. It should at least be great than 50
for normal operations.
M91
Displays the Time Ratio between the Measured Total Transit Time and the Calculated
time. If the pipe parameters are entered correctly and the transducers are properly
installed, the ratio value should be in the range of 100±3%. Otherwise the entered
parameters and the transducer installation should be checked.
M92
Displays the estimated fluid sound velocity. If this value has an obvious difference with
the actual fluid sound speed, pipe parameters entered and the transducer installation
should be checked again.
M93
Displays total transit time and delta time(transit time difference)
M94
Displays the Reynolds number and the pipe factor used by the flow rate measurement
program. Pipe factor is calculated based on the ratio of the line-average velocity and the
cross-section average velocity.
M95
(1) Display the positive and negative energy totalizers
(2) Upon entering this window, the circular display function will be started
automatically. The following windows will be displayed one by one, each window will
stay for 8 seconds: M95>>M00>>M01>>M02>>M02>>
M03>>M04>>M05>>M06>>M07>>M08>>M90>>M91>>M92>> M93>>
M94>>M95. This function allows the user to visit all the important information without
any manual action.
To stop this function, simply press a key. Or switch to a window other than M95.
M96
This is not a window but a command for the thermal printer to advance 5 lines of paper.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tsonic TUF-2000 Series and is the answer not in the manual?

Tsonic TUF-2000 Series Specifications

General IconGeneral
BrandTsonic
ModelTUF-2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals