Calibration
Calibration of the Prole device is made during production. Setting or calibration can be dis-
trupted during transportation, or if disassembling of the units requiring calibration took place.
Deviation <= 0.3 ° are not essential. It is not recommended to produce adjustments for the
elimination of non-essential deviations. All adjustments should be performed on a at, solid
surface.
Calibration of horizontal frame
Option 1.
1. Place the digital protractor on the reference plane and reset its value.
2. Place the digital protractor on the frame. If the deviation is less than 1 °, the adjustment is
not needed. Further actions are necessary to make an adjustment.
15