1.1.1
Operation and Maintenance
Manual
Electronic Laboratory Autoclaves
Models LABSCI 11LV
First time published; 06.22.2014
Current revision (G) published 01.17.2016
ELV-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
WR: water recycling system
F: optional fan for super-fast cooling
Cat. No. MAN205-0468003EN Rev G
Heidolph North America , 1235 N Mittel Blvd. Suite B Wood Dale IL 60191
Phone: 224-265-9600 Fax: 224-265-9611, e-mail: service@heidolph.com

Table of Contents

Questions and Answers

  • J
    Jasmine KelleySep 8, 2025
    What does 'Low Pressure' mean on my Tuttnauer Laboratory Equipment and how to resolve it?
    • L
      Laura MccoySep 8, 2025
      The 'Low Pressure' message on your Tuttnauer Laboratory Equipment indicates that the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • A
    Anna BarnesAug 25, 2025
    What does 'Low Pressure' mean on Tuttnauer LABSCI 11LV ELV Laboratory Equipment?
    • L
      lynn74Aug 25, 2025
      If you see 'Low Pressure' displayed on your Tuttnauer Laboratory Equipment, it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To address this, perform a new cycle.
  • H
    Harold RossAug 25, 2025
    What does 'High Temp. (Ending)' mean on Tuttnauer LABSCI 11LV ELV Laboratory Equipment?
    • J
      Joshua DouglasAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Temp. (Ending)', it means the system couldn't reach the required temperature in the chamber within 10 minutes. To resolve this, simply perform a new cycle.
  • J
    Joshua SmithSep 22, 2025
    What to do for 'Supply water error' on Tuttnauer LABSCI 11LV ELV?
    • P
      penningtonthomasSep 22, 2025
      If your Tuttnauer Laboratory Equipment displays 'Supply water error (digitat input option)', it indicates a malfunction in the city (tap) water supply. To resolve this, check and fix the city (tap) water supply.
  • W
    wolfegilbertSep 9, 2025
    What does 'High Pressure' mean on my Tuttnauer Laboratory Equipment and how can I fix it?
    • A
      andrew16Sep 9, 2025
      The 'High Pressure' message on your Tuttnauer Laboratory Equipment indicates that the chamber pressure rose 4.2 psi-29 kPa above the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • H
    Haley MaldonadoSep 14, 2025
    What does 'Air Error' mean on my Tuttnauer Laboratory Equipment and how can I fix it?
    • E
      Erica HenrySep 14, 2025
      The 'Air Error' message on your Tuttnauer Laboratory Equipment indicates that the autoclave did not reach atmospheric pressure after 100 minutes at the end of the cycle. Wait until the autoclave reaches atmospheric pressure and perform a new cycle.
  • R
    Robert Smith DVMSep 3, 2025
    What does 'Low Pressure (Cooling)' mean on my Tuttnauer Laboratory Equipment and how can I resolve it?
    • K
      Kimberly PinedaSep 3, 2025
      The 'Low Pressure (Cooling)' message on your Tuttnauer Laboratory Equipment means that the pressure in the chamber did not reach the preset pressure before the cooling stage began. Check and fix the compressed air supply.
  • D
    Dustin KeithSep 4, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows a compressed air supply error?
    • J
      Julie WeberSep 4, 2025
      If your Tuttnauer Laboratory Equipment displays a compressed air supply error, check and fix the compressed air supply.
  • S
    Sheila CabreraSep 23, 2025
    What does 'Canceled By User' mean on my Tuttnauer Laboratory Equipment?
    • L
      Lauren Villa DVMSep 23, 2025
      The 'Canceled By User' message on your Tuttnauer Laboratory Equipment indicates that the START/STOP key was pressed and the cycle was aborted. Wait until 'cycle failed – canceled by user' or 'cycle end – canceled by user' is displayed, and then perform a new cycle.
  • S
    Sharon SmithAug 29, 2025
    What does 'Low Pressure (Cooling)' mean on Tuttnauer LABSCI 11LV ELV Laboratory Equipment?
    • I
      inelsonAug 29, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Pressure (Cooling)', it means the pressure in the chamber did not reach the preset pressure before initiating the cooling stage. To resolve this, check and fix the compressed air supply.

Tuttnauer LABSCI 11LV ELV Specifications

General IconGeneral
BrandTuttnauer
ModelLABSCI 11LV ELV
CategoryLaboratory Equipment
LanguageEnglish