OPERATION
&
MAINTENANCE
MANUAL
Table-Top Autoclaves
Models 1730, 2340, 2540, 3545, 3850, 3870M

Other manuals for Tuttnauer 1730

Questions and Answers

  • C
    Caleb BaileyAug 15, 2025
    What to do if the POWER indicator light doesn't light up on my Tuttnauer 1730?
    • O
      olevineAug 15, 2025
      If the main switch is in the START position but the POWER indicator light doesn't illuminate on your Tuttnauer Laboratory Equipment, first ensure the equipment is receiving power. Check the circuit breaker and lift the lever if it has tripped. If power is available, the main switch itself may be defective and require checking or replacement. You can test the bulb by setting the timer for 15 minutes and turning the multi-purpose valve to the DRY position; if the HEAT & DRY lights illuminate, the POWER bulb is likely burnt and needs replacement. If none of the lights turn on, there may be a fault in the electrical line, which needs to be checked and repaired from the entry point to the main switch.
  • S
    Sydney MartinAug 18, 2025
    Why is the pressure safety valve activated and heating continues in STE position in my Tuttnauer Laboratory Equipment?
    • Y
      ylewisAug 19, 2025
      If your Tuttnauer Laboratory Equipment shows sufficient temperature and pressure, but the pressure safety valve is activated and heating continues while in STE position, it indicates that either the pressure switch (B10) is faulty and not stopping the heating at the required temperature, or the pressure safety valve itself is defective. In either case, the pressure switch (B10) should be fixed or replaced, and the pressure safety valve should be replaced if it's faulty.
  • O
    obergerAug 22, 2025
    What causes the temperature safety device to activate during sterilization cycles in my Tuttnauer Laboratory Equipment?
    • A
      Amanda BryantAug 22, 2025
      If, during sterilization cycles, the temperature safety device is activated due to overheating in your Tuttnauer Laboratory Equipment, it could be due to several reasons. Check if the amount of water in the chamber is insufficient and add water as needed. Also, verify that the thermostat is set to the correct temperature. A leaking multi-purpose valve, which allows water to return to the reservoir, or a leaking safety valve, or excessive leaks from the air trap jet can also cause overheating. Replace these components as necessary.
  • S
    Stacey SmithAug 28, 2025
    How to fix water not entering the chamber when the multi-purpose valve is in FILL position in Tuttnauer 1730?
    • L
      Linda BrowningAug 28, 2025
      If your Tuttnauer Laboratory Equipment's multi-purpose valve is in the FILL position but water isn't entering the chamber, there are a few potential causes. First, ensure the reservoir is filled with mineral-free water. If the reservoir is full, the multi-purpose valve or the pipe itself may be clogged. Follow the instructions in para. 11.10 to address these blockages.
  • J
    Jeffery KelleySep 1, 2025
    What to do if the HEAT light is not on in my Tuttnauer Laboratory Equipment?
    • W
      William LeachSep 1, 2025
      If the HEAT light is not on, while the multi-purpose valve is in STE position and the main switch is ON, and the Tuttnauer Laboratory Equipment is experiencing issues such as operating without heat, insufficient heating, or no heating at all, several factors could be responsible. If the bulb is burnt, replace it. If the pressure switch (B10) is set to too low a temperature, adjust it. A faulty Microswitch No.2 or Bridge No. 10 may also be the cause, requiring replacement or fixing, respectively. If there is no heating, the pressure switch (B10) or the safety thermostat might be faulty, necessitating repair or replacement.
  • A
    Anthony DavisSep 4, 2025
    Why does pressure build up very slowly in my Tuttnauer 1730 Laboratory Equipment?
    • C
      Collin ByrdSep 4, 2025
      If the pressure builds up very slowly in your Tuttnauer Laboratory Equipment, several factors could be at play. One or more heaters might be burnt out, requiring inspection and replacement. Alternatively, there might be too much water in the chamber; verify the correct water level. A leaky door gasket could also be the culprit; try tightening it, and if the leak persists, replace the gasket. Lastly, the safety thermostat might not be set to the right temperature, causing it to halt heating prematurely; calibrate the thermostat accordingly.
  • R
    Raymond FoxSep 8, 2025
    What to do if the door handle cannot be turned counter clockwise for opening Tuttnauer 1730?
    • M
      Margaret CampbellSep 8, 2025
      If the door handle of your Tuttnauer Laboratory Equipment cannot be turned counter-clockwise to open, it could be due to the door pin being set in the groove. Try slightly turning the handle in the closing direction (clockwise) and then attempt to open it again. If the problem persists, the door locking system may be stuck or the bellows damaged. Consult the “Pressure Door Lock System” section in the service section for further assistance. After opening the door, replace the bellows.
  • K
    Katie Caldwell DVMSep 12, 2025
    Why does dry operation not work on my Tuttnauer 1730?
    • K
      Kevin MilesSep 12, 2025
      If the dry operation does not work on your Tuttnauer Laboratory Equipment, it could be due to the dry microswitch MSw2 not being actuated or being faulty, in which case it should be replaced. Alternatively, the timer board might be faulty, requiring replacement as well.
  • E
    Erik JenkinsNov 24, 2025
    Why are temperature and pressure insufficient when POWER and HEAT lights are lit in my Tuttnauer 1730?
    • J
      Julie BrightNov 24, 2025
      If your Tuttnauer Laboratory Equipment's multi-purpose valve is in the STE position, the POWER and HEAT lights are on, but the temperature and pressure are not sufficient, it could be due to several reasons: 1. The water level might be lower than required, so add water. 2. The pressure switch (B10) might not be calibrated correctly, so calibrate it. 3. Steam could be escaping from the safety valve; try pulling the safety valve ring for 2 seconds, then release; if the leakage continues, replace the valve. 4. The air trap jet hole might be leaking excessively, so replace the air trap jet. 5. Steam could be escaping from the door seal; tighten the door locking bolt, and if leakage persists, replace the door seal. 6. Steam could be escaping from piping connections; locate the leakage an...

Summary

General Autoclave Information

Incoming Inspection

Inspect autoclave upon receipt for damage. Document any issues.

Technical Data and Specifications

Technical Specifications

Detailed specifications including chamber volume, tray dimensions, and weight.

Electrical Data

Information on power consumption, voltage, and current ratings.

Component Descriptions

Control Panel Components

Details on the buttons, indicators, and controls on the autoclave's panel.

Other Autoclave Components

Description and function of key parts outside the control panel.

Installation, Placement, and Leveling

Autoclave Leveling Procedure

Procedure for ensuring the autoclave is stable and correctly positioned.

Water Quantity for Cycles

Specifies the correct amount of water needed for sterilization cycles.

Water Quality Requirements

Maintenance Instructions

Preventive and Scheduled Maintenance

Routine tasks to keep the autoclave in optimal condition and minimize downtime.

Cleaning with Chamber Brite

Maintaining and Replacing Parts

Safety Tests After Repair

Electrical safety checks required after any repair or disassembly.

Replacing the Safety Valve

Instructions for replacing the critical safety valve component.

Replacing Heating Elements

Instructions for replacing the autoclave's heating components.

Replacing Multi-Purpose Valve

Procedure for replacing the main control valve assembly.

Tuttnauer 1730 Specifications

General IconGeneral
Model1730
TypeAutoclave
Number of Trays2
Chamber MaterialStainless Steel
Warranty1 year
Chamber Volume23 L
Power Supply120 V
Sterilization Temperature Range250°F (121°C)