EasyManua.ls Logo

Tuttnauer 5596 SP-1 - User Manual

Tuttnauer 5596 SP-1
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Artisan Technology Group is your source for quality
new and certied-used/pre-owned equipment
FAST SHIPPING AND
DELIVERY
TENS OF THOUSANDS OF
IN-STOCK ITEMS
EQUIPMENT DEMOS
HUNDREDS OF
MANUFACTURERS
SUPPORTED
LEASING/MONTHLY
RENTALS
ITAR CERTIFIED
SECURE ASSET SOLUTIONS
SERVICE CENTER REPAIRS
Experienced engineers and technicians on staff
at our full-service, in-house repair center
WE BUY USED EQUIPMENT
Sell your excess, underutilized, and idle used equipment
We also offer credit for buy-backs and trade-ins
www.artisantg.com/WeBuyEquipment
REMOTE INSPECTION
Remotely inspect equipment before purchasing with
our interactive website at www.instraview.com
LOOKING FOR MORE INFORMATION?
Visit us on the web at www.artisantg.com for more
information on price quotations, drivers, technical
specications, manuals, and documentation
Contact us: (888) 88-SOURCE | sales@artisantg.com | www.artisantg.com
SM
View
Instra

Questions and Answers

  • J
    jill28Aug 16, 2025
    How to troubleshoot a Tuttnauer Laboratory Equipment when the control unit is not energized?
    • R
      richardsonstephenAug 16, 2025
      If the control unit of your Tuttnauer Laboratory Equipment is not energised and there are no displays or lights on the front panel, check and switch on the circuit breaker. Alternatively, the power switches on the front panels of the autoclave or steam generator may be off or defective, in which case the defective switch should be replaced.
  • S
    stacey74Aug 19, 2025
    What to do if the steam pressure goes beyond the control point in Tuttnauer 5596 SP-1 Laboratory Equipment?
    • L
      Laura SantosAug 19, 2025
      If the steam pressure in your Tuttnauer Laboratory Equipment goes beyond the control point, the pressure switch may be faulty or incorrectly set; in this case, check, repair, or replace the pressure switch. Alternatively, the contacts at the contactor may be welded because of a short circuit; check the cause and replace the damaged contactor.
  • K
    Krista MillerAug 27, 2025
    How to fix Tuttnauer Laboratory Equipment when the drying stage does not work?
    • N
      nhuffAug 27, 2025
      If the drying stage of your Tuttnauer Laboratory Equipment does not work, refer to par. 2 and 4 for troubleshooting the vacuum system. The steam valve might be leaking, so repair or replace the steam valve. There could also be control system trouble, so replace the control boards.
  • M
    Molly FloydAug 30, 2025
    Why does temperature rise above the preset sterilization value in Tuttnauer 5596 SP-1?
    • J
      Jeremy JohnsonAug 30, 2025
      If the temperature rises above the preset sterilization value in your Tuttnauer Laboratory Equipment, it could be due to steam pressure being too high; check and correctly set the pressure switch of the generator and the pressure reducer. The steam valve may also be leaking, so fix or replace the steam valve. Alternatively, there may be control system trouble, so replace control boards.
  • J
    James AndrewsSep 2, 2025
    What causes unsatisfactory sterilization results in Tuttnauer 5596 SP-1 Laboratory Equipment?
    • T
      Tony PerezSep 2, 2025
      If you are experiencing unsatisfactory results from the sterilization process of your Tuttnauer Laboratory Equipment, refer to par. 2 and 4 for trouble related to the vacuum system, as incomplete removal of air pockets could be the cause. Also, the sterilization conditions may have improper temperature or pressure that is too low, or the sterilization time may be too short; set the temperature and time correctly according to the sort and quantity of material. Incorrect loading of the autoclave, leading to air trapped in the load, can also cause this issue; refer to the section on ‘Preparation before Sterilization’.
  • Z
    Zachary JohnsonSep 4, 2025
    What to do if exhaust phases do not function in Tuttnauer 5596 SP-1 Laboratory Equipment?
    • W
      William KennedySep 4, 2025
      If the fast or slow exhaust phases of your Tuttnauer Laboratory Equipment do not function, and the chamber pressure remains high or exhaust takes too long, the fast or slow exhaust valves may be stuck or damaged, requiring repair or replacement of the damaged valve. The steam valve may also be leaking, so repair or replace the steam valve. Alternatively, there may be control system trouble, so replace control boards.
  • J
    Justin GoodmanSep 7, 2025
    How to resolve if the air inlet does not work on Tuttnauer 5596 SP-1?
    • S
      Stephen MartinezSep 7, 2025
      If the air inlet of your Tuttnauer Laboratory Equipment does not work, the air inlet valve might be clogged or damaged, requiring repair or replacement of the air valve. Alternatively, the air filter could be clogged, so replace the air filter.
  • M
    Michele JacksonSep 12, 2025
    Why are packs remaining wet after drying in Tuttnauer 5596 SP-1?
    • E
      Eric CookSep 12, 2025
      If your Tuttnauer Laboratory Equipment is producing incomplete drying and packs remain wet, the steam pressure to the jacket may be insufficient, so set or fix the pressure switch. There may also be insufficient vacuum; refer to problems related to vacuum in par. 2 and 4. The steam trap may not be eliminating the condensate, so clean or repair the steam trap. Finally, incorrect loading of the autoclave can cause this issue, so reload the autoclave appropriately.
  • K
    Keith JonesNov 21, 2025
    Why does the steam generator not operate in Tuttnauer 5596 SP-1 Laboratory Equipment?
    • D
      deankimberlyNov 21, 2025
      If the steam generator in your Tuttnauer Laboratory Equipment does not operate, it could be due to several reasons. The water pump may be defective or blocked, requiring repair or replacement of the pump or motor. There might be a blockage or low water pressure in the pump circuit, necessitating adjustment of water pressure and ensuring free water circulation. A circuit breaker may be tripping due to overload or damage, requiring you to check and remove the overload or replace the circuit. The upper electrode could be short-circuited, needing cleaning or replacement. Electronic pressure transducers or measuring circuits might be faulty or out of calibration, so check and calibrate the pressure switches or replace them. The lower electrode may be dirty or rusted, requiring cleaning or repla...
  • F
    farmerdonaldNov 24, 2025
    What to do if the conditioning phase (pre-vacuum) doesn't work on Tuttnauer 5596 SP-1 Laboratory Equipment?
    • H
      Harold WilliamsNov 24, 2025
      If the conditioning phase (pre-vacuum) does not work on your Tuttnauer Laboratory Equipment, refer to solutions indicated in par. 2 to check if the trouble is due to pump or its electrical supply circuit. The vacuum valve passing water through the vacuum pump may be stuck or damaged, requiring you to check, repair, or replace the water to pump valve. The exhaust to pump valve may be stuck or damaged, so check, repair, or replace the fast exhaust valve. Ensure the door is properly closed and the gasket is not damaged or incorrectly seated on the groove; if necessary, close the door, fix, or replace the gasket. Check for leakage at the vacuum or exhaust piping and tighten fitting joints or replace defective packing. The steam inlet valve may be blocked or damaged, so fix or replace the steam...

Summary

General Information

Introduction to Autoclave

Overview of the autoclave's purpose, capabilities, and features.

Autoclave Standards and Certifications

Applicable standards and certifications the autoclave meets.

Autoclave Technical Specifications

Detailed technical specifications of the autoclave model.

Functional Description

Piping System Overview

Explanation of the autoclave's piping system and valve functions.

Pneumatic Control System Details

Description of the compressed air system controlling pneumatic valves.

Electric System Components

Details of the autoclave's electrical components and wiring.

Water Quality Requirements

Requirements for water used in the steam generator and vacuum system.

Sterilization Programs

Sterilization Programs Details

Details of pre-set sterilization cycles and their parameters.

Control and Monitoring

CAT 2007 Control Panel

Overview and description of the autoclave's control panel interface.

System Messages and Safety Measures

Meanings of system messages and safety interlocks.

Control Panel Operation Guide

Step-by-step guide for operating the control panel.

Calibration

Temperature and Pressure Calibration

Procedure for calibrating temperature and pressure sensors.

Preparation Before Sterilization

Loading and Unloading Procedures

Guidelines for safely loading and unloading the sterilizer chamber.

Installation Instructions

Autoclave Mounting Instructions

Procedures for physically installing the autoclave.

Utility Connections

Connecting the autoclave to steam, air, water, and electricity.

Operating Instructions

Autoclave Operation Steps

How to run a sterilization cycle.

Maintenance of the Autoclave

Preventive and Periodical Maintenance

Routine checks and cleaning schedules.

Safety Tests After Repair

Electrical safety tests to be performed after maintenance.

Troubleshooting Guide

Common issues, causes, and corrective actions.

Spare Parts

Spare Parts List

List of available spare parts and their descriptions.

Valves Numbering

Valves Numbering and Functions

Mapping of valve numbers to their functions in the system.

Pressure vs Temperature Table

Saturated Steam Data

Table correlating pressure and temperature for saturated steam.

XPCS Manual

XPCS Software Usage

Instructions for configuring and using the XPCS software.

XPCS Download Tasks

Steps for downloading software and parameters.

Tuttnauer 5596 SP-1 Specifications

General IconGeneral
TypeAutoclave
Model5596 SP-1
Voltage230V
Frequency50/60 Hz
Chamber MaterialStainless Steel
Exterior MaterialStainless Steel
Door TypeAutomatic sliding door
Safety FeaturesOverpressure protection
Trays2