What to do if my Tuttnauer Laboratory Equipment displays Low Temp?
- AAnna BaileySep 18, 2025
If your Tuttnauer Laboratory Equipment displays 'Low Temp', perform a new cycle.
What to do if my Tuttnauer Laboratory Equipment displays Low Temp?
If your Tuttnauer Laboratory Equipment displays 'Low Temp', perform a new cycle.
What to do if my Tuttnauer 5050 ELV-MD displays Compressed air supply error (digital input option)?
If your Tuttnauer Laboratory Equipment displays 'Compressed air supply error (digital input option)', check and fix the air supply.
What to do if my Tuttnauer 5050 ELV-MD displays No Water?
If your Tuttnauer Laboratory Equipment displays 'No Water': 1. Check and fix the mineral free water supply. 2. Check and clean the water inlet filter. 3. Clean the water level electrode.
What to do if my Tuttnauer Laboratory Equipment displays Compressed air supply error?
If your Tuttnauer Laboratory Equipment displays 'Compressed air supply error', check and fix the compressed air supply.
What to do if my Tuttnauer 5050 ELV-MD displays Canceled By User?
If your Tuttnauer Laboratory Equipment displays 'Canceled By User', wait until "cycle failed – canceled by user" or "cycle end – canceled by user" is displayed. Perform a new cycle.
What to do if my Tuttnauer 5050 ELV-MD Laboratory Equipment displays RTC Error - Please Set Current Date and Time?
If your Tuttnauer Laboratory Equipment displays 'RTC Error - Please Set Current Date and Time', set the current date and time. If the problem persists, further assistance is needed.
What to do if the Tuttnauer 5050 ELV-MD Laboratory Equipment displays Low Vacuum?
If your Tuttnauer Laboratory Equipment displays a 'Low Vacuum' message and the FAIL indicator lights up because a vacuum level of 25kPa wasn't reached within 20 minutes of starting the cycle, you can try the following: 1. Start a new cycle. 2. Consider that the bio-hazard filter may be clogged. If the problem persists, further assistance is needed.
Why is my Tuttnauer Laboratory Equipment leaking at the door?
If your Tuttnauer Laboratory Equipment is leaking at the door, it could be due to: 1. A dirty door gasket: Clean the door gasket. 2. A damaged door gasket: Further assistance is needed.
What to do if my Tuttnauer 5050 ELV-MD displays Heat Time Error?
If your Tuttnauer Laboratory Equipment displays 'Heat Time Error', verify that the autoclave is not overloaded.
What to do if my Tuttnauer 5050 ELV-MD Laboratory Equipment displays Low Pressure?
If your Tuttnauer Laboratory Equipment displays 'Low Pressure', perform a new cycle.
| Type | Autoclave |
|---|---|
| Sterilization Time Range | 1 to 99 minutes |
| Material | Stainless steel |
| Sterilization Temperature Range | 105°C to 135°C |
| Door Type | Automatic Sliding Door |
| Safety Features | Over-Temperature Protection |