Operation and
Maintenance Manual
Electronic Laboratory Autoclaves
Models 5050, 5075 ELV-MD/ELV-D
ELV-MD: standard autoclave (manual door)
ELV-D: standard autoclave (automatic door)
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
WR: water recycling system
F: optional fan for super-fast cooling
Cat. No. MAN205-0451001EN Rev I
Tuttnauer Europe B.V.,
Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Questions and Answers

  • E
    Elizabeth BishopSep 18, 2025
    What to do if my Tuttnauer Laboratory Equipment displays Low Temp?
    • A
      Anna BaileySep 18, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Temp', perform a new cycle.
  • A
    Allison MaxwellAug 25, 2025
    What to do if my Tuttnauer 5050 ELV-MD displays Compressed air supply error (digital input option)?
    • C
      Cindy JonesAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays 'Compressed air supply error (digital input option)', check and fix the air supply.
  • A
    Aaron RobertsAug 25, 2025
    What to do if my Tuttnauer 5050 ELV-MD displays No Water?
    • C
      Casey LambAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'No Water': 1. Check and fix the mineral free water supply. 2. Check and clean the water inlet filter. 3. Clean the water level electrode.
  • A
    Andrea HowellAug 27, 2025
    What to do if my Tuttnauer Laboratory Equipment displays Compressed air supply error?
    • D
      Dylan RamirezAug 27, 2025
      If your Tuttnauer Laboratory Equipment displays 'Compressed air supply error', check and fix the compressed air supply.
  • B
    Bruce AndersonSep 23, 2025
    What to do if my Tuttnauer 5050 ELV-MD displays Canceled By User?
    • M
      Michael HarveySep 23, 2025
      If your Tuttnauer Laboratory Equipment displays 'Canceled By User', wait until "cycle failed – canceled by user" or "cycle end – canceled by user" is displayed. Perform a new cycle.
  • M
    Matthew WadeSep 23, 2025
    What to do if my Tuttnauer 5050 ELV-MD Laboratory Equipment displays RTC Error - Please Set Current Date and Time?
    • C
      christinapittmanSep 23, 2025
      If your Tuttnauer Laboratory Equipment displays 'RTC Error - Please Set Current Date and Time', set the current date and time. If the problem persists, further assistance is needed.
  • J
    Jason BrownSep 10, 2025
    What to do if the Tuttnauer 5050 ELV-MD Laboratory Equipment displays Low Vacuum?
    • J
      Jacob DavisSep 10, 2025
      If your Tuttnauer Laboratory Equipment displays a 'Low Vacuum' message and the FAIL indicator lights up because a vacuum level of 25kPa wasn't reached within 20 minutes of starting the cycle, you can try the following: 1. Start a new cycle. 2. Consider that the bio-hazard filter may be clogged. If the problem persists, further assistance is needed.
  • A
    Adriana DavisSep 12, 2025
    Why is my Tuttnauer Laboratory Equipment leaking at the door?
    • C
      Clinton DavisSep 12, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, it could be due to: 1. A dirty door gasket: Clean the door gasket. 2. A damaged door gasket: Further assistance is needed.
  • M
    Mark JonesSep 4, 2025
    What to do if my Tuttnauer 5050 ELV-MD displays Heat Time Error?
    • A
      andersonmelvinSep 4, 2025
      If your Tuttnauer Laboratory Equipment displays 'Heat Time Error', verify that the autoclave is not overloaded.
  • L
    Linda GomezSep 1, 2025
    What to do if my Tuttnauer 5050 ELV-MD Laboratory Equipment displays Low Pressure?
    • S
      Sharon PopeSep 1, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Pressure', perform a new cycle.

Tuttnauer 5050 ELV-MD Specifications

General IconGeneral
TypeAutoclave
Sterilization Time Range1 to 99 minutes
MaterialStainless steel
Sterilization Temperature Range105°C to 135°C
Door TypeAutomatic Sliding Door
Safety FeaturesOver-Temperature Protection