What does it mean if the chamber temperature is not in range on my Tuttnauer Laboratory Equipment?
- VVirginia FordAug 30, 2025
If your Tuttnauer Laboratory Equipment displays that the chamber temperature is not in range, it could be due to several reasons: the 'End Temperature' parameter has been changed and is too low, a faulty temperature sensor, or a faulty command to the heating elements due to a short-circuit or a faulty I/O board. Correct the parameter. If the issue persists, replace the faulty sensor. Also, fix any short-circuits or replace the I/O board.