OPERATION
&
MAINTENANCE
MANUAL
Laboratory Vertical Steam Sterilizers
models 5050, 5075
ELV-D
Standard Autoclave
ELVC-D
Including Fast Cooling System
Cat. No. MAN205-0451001EN Rev B
Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510,
Fax: +31/76-5423540
Tuttnauer U.S.A. Co, Ltd. 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836,
Fax: (631) 737 0720

Table of Contents

Other manuals for Tuttnauer 5075

Questions and Answers

  • S
    Shannon LangSep 23, 2025
    What to do if Tuttnauer 5075 displays ‘Low Heat’?
    • J
      Jason GatesSep 23, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
  • J
    John PalmerAug 27, 2025
    What does 'Heat Time Error (Keep)' mean on my Tuttnauer Laboratory Equipment, and how can I fix it?
    • D
      Deborah GilbertAug 27, 2025
      If your Tuttnauer Laboratory Equipment displays a 'Heat Time Error (Keep)' message, it means the system couldn't reach the required temperature in the chamber during the optional "Keep Heat" stage within the preset time. To resolve this, verify that the autoclave is not overloaded.
  • J
    jillhernandezAug 27, 2025
    What does 'High Pressure (Exhaust)' mean on my Tuttnauer Laboratory Equipment and how do I fix it?
    • J
      Jean MitchellAug 27, 2025
      If your Tuttnauer Laboratory Equipment displays a 'High Pressure (Exhaust)' message, it means the system couldn't reach the preset pressure within 10 minutes from the beginning of the exhaust stage. This can be due to the exhaust valve not opening because of a faulty connection or a faulty I.O. board, or the exhaust valve itself might be faulty. First, perform a new cycle. If the problem persists, fix the connection, replace the faulty I.O. board, or replace the faulty valve.
  • V
    Victoria WellsSep 1, 2025
    What does High Temp error mean on Tuttnauer 5075 Laboratory Equipment and how to solve it?
    • A
      Alexander RamosSep 1, 2025
      The 'High Temp' message on your Tuttnauer Laboratory Equipment is displayed if the temperature rises 7°F (4°C) above the sterilization temperature during the sterilization stage for 2 seconds during sterilization cycle. To solve this, perform a new cycle.
  • K
    Katrina FryeSep 3, 2025
    What does High Pressure (Exhaust) error mean on Tuttnauer Laboratory Equipment and how to solve it?
    • R
      Roberto SmithSep 3, 2025
      The 'High Pressure (Exhaust)' message on your Tuttnauer Laboratory Equipment means the system couldn't reach the preset pressure within 10 minutes from the start of the exhaust stage. To address this, perform a new cycle.
  • K
    Kayla GregorySep 5, 2025
    How to fix RTC Error on Tuttnauer 5075?
    • A
      Angel FigueroaSep 5, 2025
      Set the current date and time on your Tuttnauer Laboratory Equipment. If the problem continues, call a technician.
  • C
    cassandra54Sep 15, 2025
    What should I do if my Tuttnauer Steriliser displays 'Power Down'?
    • C
      christinapriceSep 15, 2025
      If the Tuttnauer Steriliser displays 'Power Down', it means a power outage occurred during the cycle. Turn the autoclave back on, wait until it reaches a safe condition, and then perform a new cycle.
  • D
    debbiejonesSep 12, 2025
    What to do if my Tuttnauer Steriliser displays 'Low Pressure'?
    • C
      curtisnicoleSep 13, 2025
      If the Tuttnauer Steriliser displays 'Low Pressure', it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To solve this, perform a new cycle.
  • M
    Mr. Ronald JensenSep 23, 2025
    Why is Tuttnauer 5075 leaking at the door?
    • N
      ncoleSep 23, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
  • F
    fperezAug 25, 2025
    What does 'Low Temp' mean on my Tuttnauer Laboratory Equipment and how to fix it?
    • X
      xandersonAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays a 'Low Temp' message, it indicates that the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. This can also be caused by a leaking door gasket. Try performing a new cycle. If the issue continues, replace the door gasket.

Summary

OPERATION & MAINTENANCE MANUAL

Laboratory Vertical Steam Sterilizers 5050, 5075

Details the 5050 and 5075 vertical steam sterilizers.

1. GENERAL

Incoming Inspection and Warranty

Covers initial inspection of the unit and warranty terms.

2. SAFETY INSTRUCTIONS

General Safety Guidelines

Crucial safety precautions for autoclave operation, maintenance, and usage.

3. TECHNICAL DATA

Introduction and Overview

Introduces models and their purpose.

Operating Conditions and Standards

Specifies environmental, operational limits, and compliance standards.

Specifications and Electrical Data

Details technical specifications, dimensions, and electrical requirements.

Utilities and Loading Capacities

Lists required utilities and maximum loading capacities for the unit.

FRONT VIEW

Component Identification

Labels and identifies components shown in the front view diagram.

REAR VIEW

Component Identification

Labels and identifies components shown in the rear view diagram.

OVERALL DIMENSIONS

Dimensional Drawings

Provides visual representations and dimensions of the sterilizer models.

Water Quality Requirements

Built-in Safety Features

Description of Operation Stages

4. CONTROL PANEL

Control Panel Layout

Illustrates the layout of the control panel.

Display Screen Functions

Details the information displayed on the LCD screen during operation.

Keypad Operation Guide

Explains the functions and usage of the control panel keys.

Error Messages and Symbols

Lists and explains error messages and symbols for troubleshooting.

Operational Messages

Describes operational messages and their required user actions.

5. STERILIZATION PROGRAMS

Program Overview and Parameters

Summarizes available sterilization programs and their settings.

Program 1: Unwrapped Instruments

Details parameters and sequence for unwrapped instrument sterilization.

Program 2: Wrapped Instruments

Details parameters and sequence for wrapped instrument sterilization.

Program 3: Unwrapped Pouches

Details parameters and sequence for unwrapped pouch sterilization.

Program 4: Wrapped Pouches

Details parameters and sequence for wrapped pouch sterilization.

Program 5: Unwrapped Delicate Instruments

Details parameters and sequence for unwrapped delicate instrument sterilization.

Program 6: Wrapped Delicate Instruments

Details parameters and sequence for wrapped delicate instrument sterilization.

Program 7: Custom 1

Allows user configuration of parameters for custom sterilization cycles.

Program 8: Custom 2

Allows user configuration of parameters for custom sterilization cycles.

Program 9: Liquid A

Specifies parameters for sterilizing liquids in unsealed bottles.

Program 10: Liquid B

Specifies parameters for sterilizing liquids in sealed bottles.

6. PARAMETERS AND DATA MANAGEMENT

System Navigation Overview

Outlines the system's menu structure for accessing settings and data.

Table of Default Cycle Parameters

Provides a table of default parameters for custom programs.

Accessing the Main Menu

Guides users on how to navigate to the main system menu.

Custom Cycle Parameter Settings

Details how to configure parameters for Custom 1 and Custom 2 programs.

System Parameters Adjustment

Explains how to adjust system-wide settings.

Cycle History Management

Covers viewing and exporting historical cycle data.

12. SERVICE AND MAINTENANCE

Preventive Maintenance Schedule

Details scheduled tasks for maintaining the autoclave's condition.

Daily, Weekly, and Periodic Maintenance

Covers daily, weekly, and periodic cleaning and checks.

Safety Valve Checking Procedure

Procedure for testing the safety valve's functionality.

13. TROUBLESHOOTING GUIDE

Common Malfunctions and Solutions

Provides solutions for common malfunctions and error messages.

14. SPARE PARTS LIST

Available Spare Parts

Lists available spare parts with their catalog numbers.

15. ACCESSORIES

Optional Accessories

Lists optional accessories with their catalog numbers.

Tuttnauer 5075 Specifications

General IconGeneral
TypeAutoclave
Chamber Volume75 liters
Power Supply230V, 50/60Hz
Sterilization Temperature Range121°C to 134°C
Number of Trays3
Chamber MaterialStainless Steel
Weight120 kg
Power3.0 kW
Power Consumption3, 000 watts
Temperature Range105°C to 137°C
Sterilization Time Range5 to 60 minutes
Exterior MaterialSteel
Safety FeaturesOver-temperature protection, pressure relief valve
Cycle TimeVaries depending on load size and sterilization parameters