What to do if Tuttnauer 5075 displays ‘Low Heat’?
- JJason GatesSep 23, 2025
If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
What to do if Tuttnauer 5075 displays ‘Low Heat’?
If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
What does 'Heat Time Error (Keep)' mean on my Tuttnauer Laboratory Equipment, and how can I fix it?
If your Tuttnauer Laboratory Equipment displays a 'Heat Time Error (Keep)' message, it means the system couldn't reach the required temperature in the chamber during the optional "Keep Heat" stage within the preset time. To resolve this, verify that the autoclave is not overloaded.
What does 'High Pressure (Exhaust)' mean on my Tuttnauer Laboratory Equipment and how do I fix it?
If your Tuttnauer Laboratory Equipment displays a 'High Pressure (Exhaust)' message, it means the system couldn't reach the preset pressure within 10 minutes from the beginning of the exhaust stage. This can be due to the exhaust valve not opening because of a faulty connection or a faulty I.O. board, or the exhaust valve itself might be faulty. First, perform a new cycle. If the problem persists, fix the connection, replace the faulty I.O. board, or replace the faulty valve.
What does High Temp error mean on Tuttnauer 5075 Laboratory Equipment and how to solve it?
The 'High Temp' message on your Tuttnauer Laboratory Equipment is displayed if the temperature rises 7°F (4°C) above the sterilization temperature during the sterilization stage for 2 seconds during sterilization cycle. To solve this, perform a new cycle.
What does High Pressure (Exhaust) error mean on Tuttnauer Laboratory Equipment and how to solve it?
The 'High Pressure (Exhaust)' message on your Tuttnauer Laboratory Equipment means the system couldn't reach the preset pressure within 10 minutes from the start of the exhaust stage. To address this, perform a new cycle.
How to fix RTC Error on Tuttnauer 5075?
Set the current date and time on your Tuttnauer Laboratory Equipment. If the problem continues, call a technician.
What should I do if my Tuttnauer Steriliser displays 'Power Down'?
If the Tuttnauer Steriliser displays 'Power Down', it means a power outage occurred during the cycle. Turn the autoclave back on, wait until it reaches a safe condition, and then perform a new cycle.
What to do if my Tuttnauer Steriliser displays 'Low Pressure'?
If the Tuttnauer Steriliser displays 'Low Pressure', it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To solve this, perform a new cycle.
Why is Tuttnauer 5075 leaking at the door?
If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
What does 'Low Temp' mean on my Tuttnauer Laboratory Equipment and how to fix it?
If your Tuttnauer Laboratory Equipment displays a 'Low Temp' message, it indicates that the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. This can also be caused by a leaking door gasket. Try performing a new cycle. If the issue continues, replace the door gasket.
Details the 5050 and 5075 vertical steam sterilizers.
Covers initial inspection of the unit and warranty terms.
Crucial safety precautions for autoclave operation, maintenance, and usage.
Introduces models and their purpose.
Specifies environmental, operational limits, and compliance standards.
Details technical specifications, dimensions, and electrical requirements.
Lists required utilities and maximum loading capacities for the unit.
Labels and identifies components shown in the front view diagram.
Labels and identifies components shown in the rear view diagram.
Provides visual representations and dimensions of the sterilizer models.
Illustrates the layout of the control panel.
Details the information displayed on the LCD screen during operation.
Explains the functions and usage of the control panel keys.
Lists and explains error messages and symbols for troubleshooting.
Describes operational messages and their required user actions.
Summarizes available sterilization programs and their settings.
Details parameters and sequence for unwrapped instrument sterilization.
Details parameters and sequence for wrapped instrument sterilization.
Details parameters and sequence for unwrapped pouch sterilization.
Details parameters and sequence for wrapped pouch sterilization.
Details parameters and sequence for unwrapped delicate instrument sterilization.
Details parameters and sequence for wrapped delicate instrument sterilization.
Allows user configuration of parameters for custom sterilization cycles.
Allows user configuration of parameters for custom sterilization cycles.
Specifies parameters for sterilizing liquids in unsealed bottles.
Specifies parameters for sterilizing liquids in sealed bottles.
Outlines the system's menu structure for accessing settings and data.
Provides a table of default parameters for custom programs.
Guides users on how to navigate to the main system menu.
Details how to configure parameters for Custom 1 and Custom 2 programs.
Explains how to adjust system-wide settings.
Covers viewing and exporting historical cycle data.
Details scheduled tasks for maintaining the autoclave's condition.
Covers daily, weekly, and periodic cleaning and checks.
Procedure for testing the safety valve's functionality.
Provides solutions for common malfunctions and error messages.
Lists available spare parts with their catalog numbers.
Lists optional accessories with their catalog numbers.
| Type | Autoclave |
|---|---|
| Chamber Volume | 75 liters |
| Power Supply | 230V, 50/60Hz |
| Sterilization Temperature Range | 121°C to 134°C |
| Number of Trays | 3 |
| Chamber Material | Stainless Steel |
| Weight | 120 kg |
| Power | 3.0 kW |
| Power Consumption | 3, 000 watts |
| Temperature Range | 105°C to 137°C |
| Sterilization Time Range | 5 to 60 minutes |
| Exterior Material | Steel |
| Safety Features | Over-temperature protection, pressure relief valve |
| Cycle Time | Varies depending on load size and sterilization parameters |