OPERATION
&
MAINTENANCE
MANUAL
Electronic Table - Top Autoclaves
(With cooling option)
Models
2540, 3150, 3850, 3870, 5050, 5075 EL
Cat. No. MAN205-0035000EN Rev AA
Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda, The Netherlands
Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Questions and Answers

  • A
    Allison BurnsAug 17, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Water’?
    • M
      Max CunninghamAug 17, 2025
      If your Tuttnauer Laboratory Equipment displays a ‘Low Water’ message, you should: 1. Verify that there is water in the reservoir, and add water if needed. 2. Clean the water level electrode inside the vessel. 3. Check the leveling of the machine.
  • O
    ojamesAug 19, 2025
    What to do if Tuttnauer 5075 EL Laboratory Equipment displays ‘Low Pres’?
    • M
      Matthew DavisAug 20, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Pres’, verify that no instruments are in contact with the water level electrodes. Also, check the door for leakage and replace the door gasket if necessary.
  • P
    Patrick RiveraAug 22, 2025
    What to do if Tuttnauer 5075 EL Laboratory Equipment displays ‘High Pres’?
    • P
      Patrick LewisAug 22, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the water level electrode inside the vessel. Also, check the levelling of the machine.
  • D
    Dawn GriffithAug 26, 2025
    What to do if Tuttnauer 5075 EL Laboratory Equipment is not responding?
    • D
      David MurphyAug 26, 2025
      If the Tuttnauer Laboratory Equipment is not responding, you should: 1. Check the outlet and verify that the voltage requirements have been met, including phase. 2. Ensure the main switch is in the ‘On’ position. 3. Confirm the power cord is properly connected to both the machine and the mains.
  • M
    Martha BeanAug 28, 2025
    Why is Tuttnauer 5075 EL Laboratory Equipment leaking at the door?
    • J
      jennifersullivanAug 28, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
  • E
    Elizabeth PerryAug 30, 2025
    What to do if Tuttnauer 5075 EL Laboratory Equipment displays ‘High Temp’?
    • J
      Jennifer Juarez MDAug 30, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the water level electrode inside the vessel and check the leveling of the machine.
  • C
    christopherphillipsSep 2, 2025
    What to do if Tuttnauer 5075 EL Laboratory Equipment displays ‘Low Heat’?
    • R
      Rachael KleinSep 2, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
  • H
    higginslindaSep 5, 2025
    How to fix Tuttnauer 5075 EL when the printer prints but nothing is on the paper?
    • A
      Alexis HoldenSep 5, 2025
      If the printer prints, but nothing appears on the paper, ensure the paper is mounted correctly, as only one side is printable. Also, make sure the paper is inserted in the printer. If the issue persists, switch off the machine and switch it back on while pressing the feed button on the printer. If a test printout appears, the printer is working, and there may be an electronics issue.
  • C
    Cesar CruzSep 7, 2025
    Why Tuttnauer Laboratory Equipment water does not exit chamber?
    • R
      Rodney EdwardsSep 7, 2025
      If water does not exit the chamber of your Tuttnauer Laboratory Equipment due to a clogged outlet strainer, clean the strainer according to the instructions.
  • M
    mzavalaSep 10, 2025
    What to do if Tuttnauer 5075 EL printer does not print?
    • W
      Wesley SmithSep 10, 2025
      If the printer does not print, switch off the Tuttnauer Laboratory Equipment and switch it back on while pressing the feed button on the printer. If the printer still does not print the test printout, there is a problem with the printer.

Summary

SAFETY INSTRUCTIONS

General Safety Precautions

Provides essential safety instructions, user training, procedures, and handling of hazardous materials, starting from page 8.

Technical Specifications EL

Water Quality Requirements

STERILIZATION PROGRAMS

Program 1 - Unwrapped Instruments

Details on Program 1 for sterilizing unwrapped instruments at 134°C with fast exhaust.

Program 2 (Instruments)

Details on Program 2 for instruments at 121°C with fast exhaust and drying.

Program 3 (Waste)

Details on Program 3 for waste materials at 121°C with slow exhaust and drying.

Program 4 (Liquids)

Details on Program 4 for liquids at 121°C with slow exhaust, no drying.

Program 5 (Liquids)

Details on Program 5 for liquids at 105°C with slow exhaust, no drying.

Program 6 (Liquids + Cooling)

Details on Program 6 for materials with cooling at 121°C, no drying.

KEYBOARD AND DISPLAY

Front Panel Keyboard Functions

Start;Stop Key

Functions of the Start/Stop key: starting, stopping, and canceling fail messages.

Display Description

Displayed Messages and Safety Measures

Temperature Display Messages

Messages for Low Temp, Low Heat, and High Temp conditions during operation.

Pressure Display Messages

Messages for Low Pres. and High Pres. conditions during operation.

PREPARATION BEFORE STERILIZATION

General Preparation and Instrument Cleaning

Guidelines for packaging, cleaning instruments, and loading materials correctly into the chamber.

Filling the Water Reservoir

OPERATING INSTRUCTIONS

Autoclave Operation Steps

Step-by-step guide for operating the autoclave, including filling, selecting programs, loading, and starting cycles.

Loading and Unloading the Device

Safety Guidelines for Loading;Unloading

General safety precautions and protective equipment required for loading and unloading the autoclave.

Loading Procedures

Importance of correct loading for efficient air removal, steam penetration, and preventing damage.

Unloading Procedures

Procedures for safely unloading sterilized items, visual inspection, and rejection criteria.

MAINTENANCE INSTRUCTIONS

Preventive and Scheduled Maintenance

Overview of periodic maintenance to keep the device in good condition and reduce breakdown time.

Reservoir and Strainer Maintenance

Draining the Reservoir

Step-by-step guide for draining the water reservoir, including safety precautions.

Cleaning the Water Strainer

Procedure for cleaning the water strainer, including safety cautions.

Safety Valve and Door Gasket

Checking the Safety Valve

Procedure for checking the safety valve for blockage on different models.

Replacing the Door Gasket

Instructions for replacing the door gasket on specific autoclave models.

Door Safety Systems

Autoclave Cleaning

TROUBLESHOOTING

Machine Not Responding

Troubleshooting steps when the autoclave machine is not responding to commands.

Tuttnauer 5075 EL Specifications

General IconGeneral
Model5075EL
Chamber Volume75 liters
Chamber MaterialStainless Steel
Sterilization Temperature Range105°C - 138°C
Power Supply50/60Hz
Standard AccessoriesTray