OPERATION
&
MAINTENANCE
MANUAL
Electronic Table - Top Autoclaves
(With cooling option)
Models
2540, 3150, 3850, 3870, 5050, 5075 EL
Cat. No. MAN205-0035000EN Rev AA
Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda, The Netherlands
Tel: 31 (0) 765423510, Fax: 31 (0) 765423540
All manuals and user guides at all-guides.com

Table of Contents

Questions and Answers

  • D
    David BurkeAug 17, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Water’?
    • L
      Lauren GutierrezAug 17, 2025
      If your Tuttnauer Laboratory Equipment displays a ‘Low Water’ message, you should: 1. Verify that there is water in the reservoir, and add water if needed. 2. Clean the water level electrode inside the vessel. 3. Check the leveling of the machine.
  • S
    Sean WalshAug 20, 2025
    What to do if Tuttnauer 3870 EL displays ‘Low Pres’?
    • A
      Amy GarciaAug 20, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Pres’, verify that no instruments are in contact with the water level electrodes. Also, check the door for leakage and replace the door gasket if necessary.
  • R
    Richard WeberAug 22, 2025
    What to do if Tuttnauer 3870 EL displays ‘High Pres’?
    • D
      deanbrentAug 23, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the water level electrode inside the vessel. Also, check the levelling of the machine.
  • M
    marymartinAug 26, 2025
    What to do if Tuttnauer Laboratory Equipment is not responding?
    • A
      Autumn JacksonAug 26, 2025
      If the Tuttnauer Laboratory Equipment is not responding, you should: 1. Check the outlet and verify that the voltage requirements have been met, including phase. 2. Ensure the main switch is in the ‘On’ position. 3. Confirm the power cord is properly connected to both the machine and the mains.
  • O
    Olivia MartinAug 28, 2025
    Why is Tuttnauer 3870 EL Laboratory Equipment leaking at the door?
    • R
      Richard WrightAug 28, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
  • S
    Sophia RiveraAug 31, 2025
    What to do if Tuttnauer 3870 EL displays ‘High Temp’?
    • M
      Melissa JonesAug 31, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the water level electrode inside the vessel and check the leveling of the machine.
  • L
    Lauren GreenSep 2, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Heat’?
    • B
      Brian MelendezSep 2, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
  • K
    Kendra SweeneySep 5, 2025
    How to fix Tuttnauer Laboratory Equipment when the printer prints but nothing is on the paper?
    • J
      Jessica VillarrealSep 5, 2025
      If the printer prints, but nothing appears on the paper, ensure the paper is mounted correctly, as only one side is printable. Also, make sure the paper is inserted in the printer. If the issue persists, switch off the machine and switch it back on while pressing the feed button on the printer. If a test printout appears, the printer is working, and there may be an electronics issue.
  • M
    Mason OconnorSep 8, 2025
    Why Tuttnauer Laboratory Equipment water does not exit chamber?
    • X
      Xavier GilesSep 8, 2025
      If water does not exit the chamber of your Tuttnauer Laboratory Equipment due to a clogged outlet strainer, clean the strainer according to the instructions.
  • T
    Tanya JonesSep 12, 2025
    What to do if Tuttnauer 3870 EL Laboratory Equipment printer does not print?
    • D
      Daniel StewartSep 12, 2025
      If the printer does not print, switch off the Tuttnauer Laboratory Equipment and switch it back on while pressing the feed button on the printer. If the printer still does not print the test printout, there is a problem with the printer.

Tuttnauer 3870 EL Specifications

General IconGeneral
Door TypeManual
TypeAutoclave
Power1.5 kW
Programs4 Pre-set cycles
Safety FeaturesOver-temperature protection