Questions and Answers

  • T
    tylerstewartAug 20, 2025
    What does 'Air Error' mean on Tuttnauer 3840 EL-D?
    • A
      alyssa79Aug 20, 2025
      The 'Air Error' message on your Tuttnauer Laboratory Equipment appears at the end of the cycle if the autoclave does not reach atmospheric pressure after 10 minutes. To resolve this, wait until the autoclave reaches atmospheric pressure and then perform a new cycle.
  • S
    Susan ThomasAug 22, 2025
    What to do if Tuttnauer 3840 EL-D displays 'Cycle Failed'?
    • J
      justinortizAug 22, 2025
      If your Tuttnauer Laboratory Equipment displays 'Cycle Failed', perform a new cycle.
  • M
    Mark MoralesAug 25, 2025
    What does 'RTC Error - Please Set Current Date and Time' mean on Tuttnauer Laboratory Equipment?
    • K
      kelly71Aug 25, 2025
      If your Tuttnauer Laboratory Equipment displays 'RTC Error - Please Set Current Date and Time', set the current date and time. If the problem persists, it may require technical assistance.
  • C
    Christopher HurleyAug 26, 2025
    What to do if Tuttnauer Laboratory Equipment says 'Door is open (During the cycle)'?
    • J
      Julia DavisAug 26, 2025
      If your Tuttnauer Laboratory Equipment shows 'Door is open (During the cycle)', close the door to perform a new cycle.
  • T
    Timothy BrownAug 27, 2025
    How to fix a Tuttnauer 3840 EL-D Laboratory Equipment machine that is not responding?
    • T
      Taylor ChapmanAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine is not responding, check the following: 1. Ensure the main switch is in the ‘On’ position. 2. Verify the power cord is properly connected to both the machine and the mains. 3. Check if the circuit breaker has tripped; if so, lift the circuit breaker lever.
  • R
    russell75Aug 29, 2025
    What to do if Tuttnauer Laboratory Equipment shows a 'No Water' message?
    • M
      matthew37Aug 29, 2025
      If your Tuttnauer Laboratory Equipment displays a 'No Water' message, it means the electrode in the chamber didn't detect water in time. To resolve this: 1. Check and fix the mineral-free water supply. 2. Check and clean the water inlet filter. 3. Clean the water level electrode.
  • L
    Lisa ConwayAug 31, 2025
    What does 'Low Vacuum' mean on Tuttnauer 3840 EL-D Laboratory Equipment?
    • K
      Kevin FernandezAug 31, 2025
      The 'Low Vacuum' message on your Tuttnauer Laboratory Equipment indicates that a vacuum level of 15kPa was not reached within 20 minutes after the cycle started during the air removal stage. To address this: 1. Try performing a new cycle. 2. If the issue persists, it may require technical assistance. 3. The bio-hazard filter may be clogged, requiring technical service since the door cannot be opened until a successful cycle is completed.
  • M
    Michelle SolisSep 2, 2025
    Why is my Tuttnauer 3840 EL-D printer not printing?
    • M
      Michael JohnsonSep 2, 2025
      If the printer on your Tuttnauer Laboratory Equipment is not printing, it could be due to the following reasons: 1. The printer might be out of paper. 2. As a general solution, try switching the machine off and then back on to restart it.
  • W
    walkermarissaSep 4, 2025
    What does 'High Temp' mean on Tuttnauer Laboratory Equipment?
    • R
      Robert JonesSep 4, 2025
      The 'High Temp' message on your Tuttnauer Laboratory Equipment indicates that the temperature rose 7°F (4°C) above the sterilization temperature for 2 seconds during the sterilization cycle. To resolve this, perform a new cycle.
  • A
    Alexis YuSep 6, 2025
    What does 'Low Temp' mean on Tuttnauer Laboratory Equipment?
    • D
      Diana OrtizSep 6, 2025
      The 'Low Temp' message on your Tuttnauer Laboratory Equipment means that the temperature dropped for more than 1 second below the sterilization temperature during the sterilization cycle. To fix this, perform a new cycle.