OPERATION
&
MAINTENANCE
MANUAL
Cat. No. MAN205-0458001EN Rev E
Tuttnauer Europe b.v., Hoeksteen 11, 4815 PR, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands.
+31/76-5423510,
Fax: +31/76-5423540
Model 3870 HSG-D

Table of Contents

Questions and Answers

  • J
    Jennifer MooreAug 20, 2025
    What to do if Tuttnauer 3870 HSG-D Laboratory Equipment shows "Heat Time Error"?
    • D
      Deborah HenryAug 20, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Heat Time Error", it means the system couldn't reach the required temperature in the chamber within the preset time. To resolve this, check that the autoclave is not overloaded.
  • L
    Leah WagnerAug 22, 2025
    What to do if Tuttnauer 3870 HSG-D displays "Pressure Time Error"?
    • K
      Kiara HamiltonAug 22, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Pressure Time Error", this indicates that the system could not reach the required pressure conditions in the chamber after the preset time during the air removal stage. Ensure that the autoclave is not overloaded.
  • B
    beardnicoleAug 25, 2025
    What to do if Tuttnauer 3870 HSG-D shows "Low Temp" message?
    • T
      Tony GillAug 25, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Low Temp", it means the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. Performing a new cycle may resolve this issue.
  • S
    Summer YoungAug 26, 2025
    What to do if Tuttnauer Laboratory Equipment shows "High Temp" message?
    • J
      jamesdanielAug 26, 2025
      If the Tuttnauer Laboratory Equipment displays the message "High Temp", it means the temperature rose 7°F (4°C) above the sterilization temperature during the sterilization stage for 2 seconds during the sterilization cycle. To fix it, perform a new cycle.
  • F
    fischerstevenAug 28, 2025
    What to do if Tuttnauer 3870 HSG-D displays "Low Pressure"?
    • P
      pachecojenniferAug 28, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Low Pressure", it indicates that the pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. You should perform a new cycle.
  • J
    Jose JacksonAug 30, 2025
    What to do if Tuttnauer 3870 HSG-D Laboratory Equipment displays "High Pressure"?
    • J
      Jared SchultzAug 30, 2025
      If the Tuttnauer Laboratory Equipment displays the message "High Pressure", it means the chamber pressure raised 4.2 psi-29 kPa above the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To solve this, perform a new cycle.
  • R
    Ruben WilsonAug 31, 2025
    What to do if Tuttnauer 3870 HSG-D Laboratory Equipment displays "Air Error" at the end of the cycle?
    • S
      Shawn McneilAug 31, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Air Error" at the end of the cycle, it means that the autoclave did not reach atmospheric pressure after 10 minutes. Wait until the autoclave reaches atmospheric pressure, and then perform a new cycle.
  • R
    Rebecca WallaceSep 2, 2025
    What to do if Tuttnauer Laboratory Equipment displays "Door is open (During the cycle)"?
    • J
      Jacob WileySep 2, 2025
      If your Tuttnauer Laboratory Equipment displays "Door is open (During the cycle)", this means the door was opened during the cycle. Close the door and start a new cycle.
  • M
    Melanie PonceSep 4, 2025
    What does "Power Down" message mean on Tuttnauer 3870 HSG-D?
    • K
      Kayla JohnstonSep 4, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Power Down", it indicates that a power outage occurred during the cycle. Turn the autoclave back on, wait until it reaches a safe condition, and then perform a new cycle.
  • B
    bmillerSep 6, 2025
    What does "Cycle Failed" mean on Tuttnauer 3870 HSG-D?
    • H
      Hunter RowlandSep 6, 2025
      If the Tuttnauer Laboratory Equipment displays "Cycle Failed" with a specific symbol, it indicates that an error occurred before the sterilization cycle was completed. You should perform a new cycle.

Summary

GENERAL

Incoming Inspection

Procedure for checking the autoclave for damage upon receipt.

Warranty Information

Details on instrument's warranty coverage and limitations.

Warranty Statement and Service

Warranty registration and contact info for technical service.

Safety Instructions and Operation

Safety Instructions

General safety guidelines and precautions for autoclave operation.

Operating Safety Procedures

Specific safety measures and procedures for autoclave users.

Autoclave Safety Features

Description of built-in safety mechanisms for hazard protection.

TECHNICAL DATA

Introduction to Technical Data

Overview of the autoclave's design, purpose, and capabilities.

Operating Conditions and Cautions

Specifies environmental requirements and operational cautions.

Construction Materials

Details the materials used in the autoclave's construction.

Environment Emission Information

Provides data on sound level and heat transmission.

Directives and Standards Compliance

Lists relevant directives and standards the autoclave complies with.

Technical Specifications Overview

Lists detailed specifications like chamber size, weight, and load capacity.

Electrical Data Specifications

Outlines the electrical properties and requirements of the autoclave.

Utility Requirements

Details water inlet, connections, steam outlet, and drain requirements.

Sticker Symbol Description

Explains the meaning of various symbols and warning labels on the autoclave.

Water Quality Requirements

Specifies the required quality of distilled or mineral-free water.

CONTROL PANEL

Front Panel Keyboard Overview

Details the display screen, keypad, and printer functions.

Keypad Operation Details

Explains the functions of the UP, DOWN, and START/STOP keys.

Printer Functionality

Describes the printer's role in recording cycle information.

Error Message Interpretation

Categorizes failures and refers to troubleshooting for error messages.

Operational Messages and Symbols

Explains common operational messages and required user actions.

STERILIZATION PROGRAMS

Sterilization Program Overview

Lists and describes the five sterilization programs and two test programs.

Program 1: Unwrapped Instruments

Details parameters and sequence for unwrapped instruments.

Program 2: Wrapped Instruments

Details parameters and sequence for wrapped instruments.

Program 3: Wrapped Pouches

Details parameters and sequence for wrapped pouches.

Program 4: Unwrapped Delicate Instruments

Details parameters and sequence for unwrapped delicate items.

Program 5: Wrapped Delicate Instruments

Details parameters and sequence for wrapped delicate items.

Test 1: Leak Test Procedure

Describes the procedure and purpose of the leak test.

Test 2: Bowie & Dick Test

Details the Bowie & Dick test parameters and sequence.

SCREENS DISPLAY

Successful Cycle Screens

Displays various screens shown during a successful sterilization cycle.

Aborted Cycle Screen Displays

Shows screens for cycles aborted after sterilization stage.

Failed Cycle Screen Displays

Displays screens indicating a cycle failure or cancellation.

PRINTER OPERATIONS

Printer Output Information

Explains the information printed by the autoclave during cycles.

Printer Handling and Setup

Instructions for maintaining and setting up the printer.

Loading Printer Paper

Step-by-step guide for loading paper into the printer.

Thermal Paper Storage Guidelines

Guidelines for storing and handling thermal printer paper.

PREPARATION FOR STERILIZATION

Item Preparation Principles

Outlines principles for packaging and preparing items for sterilization.

Instrument Preparation Steps

Specific instructions for preparing various types of instruments.

Pack Preparation Guidelines

Guidelines for packaging instruments and materials into packs.

Tubing Preparation Instructions

Instructions for preparing tubing for sterilization.

OPERATING INSTRUCTIONS

Autoclave Operation Initiation

Initial steps for operating the autoclave, including turning it on.

Door Opening Procedure

Step-by-step instructions for safely opening the autoclave door.

Autoclave Loading Procedure

Guidance on how to properly load items into the autoclave chamber.

Sterilization Program Selection

Instructions on choosing the appropriate sterilization program.

Door Closing Procedure

Procedure for safely and correctly closing the autoclave door.

Starting the Sterilization Cycle

How to initiate the sterilization cycle using the START/STOP key.

Autoclave Unloading Procedure

Steps for safely removing the load after the sterilization cycle.

Stopping Process and Cancelling Errors

Procedures for stopping an ongoing cycle or clearing error messages.

SERVICE AND MAINTENANCE

Preventive and Scheduled Maintenance

Overview of maintenance tasks to keep the device in good condition.

Daily Maintenance Tasks

Routine daily cleaning and checks for the autoclave.

Weekly Maintenance Tasks

Tasks to be performed weekly, including cleaning.

Periodical Maintenance Tasks

Maintenance tasks to be done periodically, including filter replacement.

Periodical Safety Tests

Monthly activation of the safety valve as a test.

Cleaning the Water Strainer

Step-by-step instructions for cleaning the water strainer.

Safety Valve Checking Procedure

Procedures for checking the chamber and generator safety valves.

PED-Approved Safety Valve Testing

Instructions for testing PED-approved safety valves.

ASME-Approved Safety Valve Testing

Instructions for testing ASME-approved safety valves.

Air Filter Replacement Guide

Step-by-step guide for replacing the bacteriological air filter.

TROUBLESHOOTING GUIDE

Machine Responsiveness Issues

Steps to take if the machine is not responding to commands.

Printer Malfunction Solutions

Solutions for printer not printing or printing blank paper.

Leak and Water Outlet Problems

Addresses problems with door leaks and water not exiting the chamber.

Temperature Error Resolution

Troubleshooting for 'Low Temp' and 'High Temp' messages.

Pressure, Time, and Heat Errors

Solutions for pressure, time, and heat time errors.

Date;Time and Door Status Errors

Addresses issues with date/time settings and door status errors.

Cycle, Air, I;O, and Power Errors

Troubleshooting for cycle failures, air errors, I/O card, and power down.

SPARE PARTS LISTING

ACCESSORIES LISTING

Tuttnauer 3870 HSG-D Specifications

General IconGeneral
BrandTuttnauer
Model3870 HSG-D
CategoryLaboratory Equipment
LanguageEnglish