Tuttnauer 3850 ELV - User Manual

Tuttnauer 3850 ELV
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Operation &
Maintenance
Manual
Laboratory Vertical Steam Sterilizers
Including Fast Cooling System
models 3840, 3850, 3870
ELV/WR/C/PV/BH/F-D
Cat. No. MAN205-0448002EN Rev F
Tuttnauer Europe b.v., Hoeksteen 11, 4815 PR, Breda, P.O. Box 7191, 4800 GD
Breda, Netherlands. +31/76-5423510, Fax: +31/76-5423540

Table of Contents

Questions and Answers

  • N
    Nathan ChambersAug 19, 2025
    What to do if 'Low Vacuum' is displayed on Tuttnauer Laboratory Equipment?
    • C
      charleslucasAug 19, 2025
      If the Tuttnauer Laboratory Equipment displays 'Low Vacuum', this indicates that a vacuum level of 15kPa was not reached within 20 minutes after the cycle started during the air removal stage. To resolve this, try performing a new cycle. If the issue persists, it may be necessary to call a technician, or the bio-hazard filter may be clogged.
  • V
    Vanessa SotoAug 21, 2025
    What does 'Supplied Distilled Water Error' mean on Tuttnauer 3850 ELV Laboratory Equipment?
    • Z
      Zachary WheelerAug 21, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Supplied Distilled Water Error", it means the mineral-free water reservoir is empty. To fix this, simply fill the mineral-free water reservoir.
  • S
    Suzanne AguilarAug 23, 2025
    What to do if Tuttnauer Laboratory Equipment displays 'No Water'?
    • J
      Justin WrightAug 25, 2025
      If the Tuttnauer Laboratory Equipment displays 'No Water', first check and fix the mineral-free water supply. Then, inspect and clean the water inlet filter, and finally, clean the water level electrode.
  • W
    William DunnAug 26, 2025
    What to do when Tuttnauer Laboratory Equipment displays 'Cycle counter exceeded -'?
    • B
      bryan97Aug 26, 2025
      If the Tuttnauer Laboratory Equipment displays 'Cycle counter exceeded -', call for service. This is because the number of cycles since the last periodical maintenance has exceeded the preset number.
  • T
    thompsondanielleAug 28, 2025
    What to do if Tuttnauer Laboratory Equipment displays 'Cycle Failed'?
    • K
      Kelly CannonAug 28, 2025
      If the Tuttnauer Laboratory Equipment displays 'Cycle Failed', perform a new cycle.
  • L
    lindseydustinAug 30, 2025
    How to troubleshoot a Tuttnauer 3850 ELV Laboratory Equipment when the display is not activated?
    • C
      cherylhigginsAug 30, 2025
      If the display of your Tuttnauer Laboratory Equipment is not activated, first, check the power source to ensure it meets the required specifications. Next, verify that the ON/OFF switch is in the ‘On’ position. Also, make sure the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and reset it if necessary. Finally, check if the circuit breaker has tripped and reset it if needed.
  • N
    Nicholas CraigSep 1, 2025
    What to do if my Tuttnauer 3850 ELV Laboratory Equipment is leaking at the door?
    • A
      andrewhuangSep 1, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure that the door is tightened sufficiently. If the leakage persists, consider replacing the door gasket.
  • A
    Ashley SloanSep 3, 2025
    How to troubleshoot a Tuttnauer 3850 ELV Laboratory Equipment machine that is not responding?
    • P
      pjonesSep 3, 2025
      If the Tuttnauer Laboratory Equipment machine isn't responding, first ensure the main switch is in the ‘On’ position. Then, check that the power cord is properly connected to both the machine and the mains. Finally, lift the circuit breaker lever if it has tripped. These issues may be caused by the main switch being off, a disconnected power cord, or a tripped circuit breaker.
  • A
    alexandrahillSep 5, 2025
    What to do if my Tuttnauer 3850 ELV printer does not print?
    • E
      Erin LongSep 5, 2025
      If the Tuttnauer Laboratory Equipment printer isn't printing, make sure the paper roll is correctly inserted. If there's still no printing, switch the machine off and then back on to restart it. This could be happening because no paper is inserted or for no obvious reason.
  • S
    Steven HughesSep 7, 2025
    Why is my Tuttnauer Laboratory Equipment leaking at the door?
    • M
      Martha ClarkSep 7, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, the door gasket may be dirty or damaged. Try cleaning the door gasket. If the leaking continues, call for service.

Summary

GENERAL

SAFETY INSTRUCTIONS

Operational Safety Guidelines

Key safety guidelines for operation, personnel, and materials.

TECHNICAL DATA

Standards, Electrical, and Specifications

Details standards, electrical data, and technical specifications.

OPERATION AND CONTROL

Description of Operation Stages

Explains the operational stages like air removal, heat, and sterilization.

Control Panel and User Interface

Details the control panel components and their functions.

Sterilization Programs

Lists and describes the available sterilization programs.

PREPARATION AND LOADING

AUTOCLAVE OPERATION

Cycle Execution and Unloading

Procedures for opening, closing, starting, and unloading the cycle.

DOOR SAFETY SYSTEM

SERVICE AND MAINTENANCE

Preventive Maintenance Schedule

Routine maintenance tasks for daily, weekly, and periodic checks.

TROUBLESHOOTING

Tuttnauer 3850 ELV Specifications

General IconGeneral
TypeAutoclave
Model3850 ELV
Number of Trays3
Frequency50/60 Hz
Power Consumption3 kW
Weight150 kg
Control SystemMicroprocessor
DisplayLCD
Printing OptionOptional
Power3000W
Voltage220-240V
Temperature Range105 to 137 °C
Operating Pressure2.1 bar
Power SupplySingle Phase
Water SupplyAutomatic
DrainageAutomatic
Sterilization Pressure Range0.22-0.35 MPa
Chamber Volume85 liters